In an axial body section, photons from a source are attenuated in tissue. In SPECT the fraction of single photons reaching the camera falls with source depth; in PET the chance that both photons escape does not depend on depth, only on total thickness. The last step compares uncorrected and Chang- or CT-corrected images and profiles of a uniform phantom.
Key point: in SPECT attenuation depends on source depth (deep structures look faint); in PET the attenuation of the photon pair depends only on total thickness, so a transmission (CT) measurement corrects it exactly and correction is mandatory.
Source: book (soft-tissue µ ≈ 0.15 cm⁻¹ at 140 keV and ≈ 0.096 at 511 keV; deep structures low, centre-to-periphery gradient; pair attenuation in PET depends on total thickness; Chang and CT-based correction); NIST XCOM water 0.154 / 0.0960 cm⁻¹. Interaction points are sampled from an exponential distribution. Phantom: 20 cm uniform disc, 64 × 64 (4 mm), 128 angles, ramp-filtered FBP, first-order Chang. Patient thickness 20 cm illustrative; scattered photons that are still detected are not shown. Atomic scale: interaction fractions in water from NIST XCOM (140.5 keV: Compton 97.6%, coherent 1.8%, photoelectric 0.6%; 511 keV: Compton 99.8%); angles from the Klein–Nishina distribution; the 435–650 keV PET window is a typical LSO/LYSO setting. The atom is not drawn to scale.
How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the source depth; ← → change steps.
Attenuation: SPECT and PET · Open in Nuclear Medicine Explained ↗