In a negative-ion cyclotron seen from above, an H⁻ ion is accelerated by RF between two dees and spirals in the magnetic field; on the right the radius grows with energy. In the second step a carbon foil strips the electrons and the proton curves the other way to the target. The third step shows the ¹⁸O(p,n)¹⁸F reaction and the F-18 saturation curve against irradiation time.
Step 1/3
Spiral acceleration
H⁻ ions from the ion source are accelerated by the RF voltage in the gap between two hollow electrodes (dees). A perpendicular magnetic field keeps them on a circular path; at every gap crossing the energy, and therefore the radius, increases: a spiral. The polarity of the dees flips every half-turn, so the ion always crosses towards the attracting dee. The orbit time does not depend on energy (cyclotron frequency).
Number of turns reduced in the drawing; in reality hundreds.
Key point: in a cyclotron a charged particle circling in a magnetic field is accelerated by RF at each crossing and spirals outwards; in a negative-ion cyclotron a carbon foil turns H⁻ into a proton to extract the beam; the products are proton rich (β⁺, EC).
Source: book (dees, RF, perpendicular magnetic field, spiral; H⁻ and carbon foil; 10-18 MeV; ¹⁸O(p,n)¹⁸F). Calculation: r = pc/(299.79·B), B = 1.4 T illustrative (44 cm at 18 MeV); cyclotron frequency qB/(2πm) ≈ 21 MHz; Q ≈ −2.44 MeV, threshold ≈ 2.57 MeV; saturation 1 − 2^(−t/109.8 min). Number of turns and energy gain illustrative; isochronous field shaping not shown.
How to use: a single click on the drawing pauses and resumes; in the second step ↑ ↓ or the buttons below change the extraction energy, in the third step the irradiation time; ← → change steps.
Negative-ion cyclotron · Open in Nuclear Medicine Explained ↗