In a narrow beam the number of photons passing falls exponentially with lead thickness; HVL and TVL, HVL by energy, and narrow- and broad-beam (build-up) geometry are shown.
Step 1/4
Exponential attenuation, HVL and TVL
A collimated Tc-99m (140 keV) beam passes through lead. Each photon has the same probability of interacting in every slice of lead, so the transmitted fraction falls exponentially: I = I₀·e^(−μx). In lead μ = 26.9 cm⁻¹ and the half-value layer HVL = 0.693/μ = 0.26 mm: each HVL halves the transmitted photons (50%, 25%, 12.5%, 6.25%). The thickness that reduces it to one tenth is TVL = 2.303/μ = 3.32 HVL.
In step 1 drag on the graph to change the thickness; in step 2 choose the radionuclide from the list on the right and the thickness below the drawing; in steps 3-4 drag on the graph to change the thickness.
Key point: each HVL halves the photons passing; HVL grows with photon energy; in a broad beam scattered photons are also measured, so transmission is higher than in a narrow beam.
Source: μ NIST XCOM (Pb 11.35 g/cm³); build-up ANS-6.4.3 (1991), isotropic point source, infinite medium.
How to use: Space or the ⏸/▶ below pauses and resumes; ↑ ↓ change the lead thickness (a short press is one notch, holding changes it smoothly); you can also drag on the graph; in the second step the buttons choose the radionuclide and thickness; ← → change steps.
HVL, TVL and the build-up factor · Open in Nuclear Medicine Explained ↗