A section through the centre of a uniform sphere, blurred by the system FWHM, is shown side by side for a hot (4:1) and a cold (0) lesion; on the right the profiles are compared with the true values. The second step shows the recovery coefficient (maximum and mean) against lesion diameter / FWHM and where the chosen lesion lies.

Step 1/2

Spill-out and spill-in

The system blurs every point to the width of its FWHM. Activity of a small hot lesion spreads outwards (spill-out) and is underestimated; activity from the surroundings spills into a small cold lesion (spill-in), which looks higher than it is. In both cases contrast falls. Make the lesion smaller: at 8 mm FWHM the centre of a 5 mm hot lesion stays far below the true value.

Hot and cold lesionhot lesion (4:1)centre 2.38 (true 4)RC_max 46%cold lesion (0:1)centre 0.54 (true 0)spill-inD = 10 mm · FWHM = 8 mm · D/FWHM = 1.3
Profile through the centrehot lesion (4:1)14cold lesion (0:1)01−20−1001020position (mm)truemeasured
Continuous stream · no seeking

Key point: when the lesion is smaller than 2-3 × FWHM a hot lesion is underestimated (spill-out) and a cold lesion overestimated (spill-in); the loss is described by the recovery coefficient and can be corrected.

Source: book (PVE < 2-3 × FWHM, spill-out / spill-in, RC = measured/true, corrected = measured/RC). Calculation: convolution of a uniform sphere with a 3D Gaussian (analytic formula; σ = FWHM/2.355); hot lesion 4:1, cold 0:1; mean RC over the sphere volume. No noise or voxel size effects.

How to use: ↑ ↓ or the buttons below change the lesion diameter; the second row picks the resolution (FWHM); ← → change steps.