An axial section shows two rows of detector rings with the patient between; annihilation photons leave at random axial angles. In 2D the septa between rings stop oblique and scattered photons; in 3D coincidences between all rings are accepted.
Step 1/2
2D (septa)
Lead or tungsten septa sit between the rings. Photons travelling obliquely hit a septum and are absorbed; only pairs reaching the same or an adjacent ring are recorded. Most photons scattered in the patient also stop in the septa: scatter and randoms are low, but because most photons are lost, sensitivity is low.
Key point: removing the septa lets cross-ring coincidences count and raises sensitivity ~4–8×; the price is more scatter and randoms and a more complex reconstruction.
Source: book (2D: same/adjacent ring, scatter ~15–20%; 3D: sensitivity ~4–8×, scatter ~30–40%, randoms↑). The drawing is schematic: the axial direction is stretched, so the ratio of counts in the drawing does not show the real gain. Photon directions isotropic; single scatter of ~30% of photons in the body is illustrative.
How to use: a single click on the drawing pauses and resumes; ← → switch between 2D and 3D.
2D and 3D PET acquisition · Open in Nuclear Medicine Explained ↗