A positron from a β⁺ emitter slows down in tissue with its real energy and speed on a common scale; it collides with tissue electrons, thermalises and annihilates; positronium formation (para and ortho) and the emission of two 511 keV photons in nearly opposite directions are shown.
Key point: positron range and the slight deviation of annihilation photons from 180° are physical limits of PET resolution.
Source: β⁺ branches and mean energies ENSDF (IAEA LiveChart); energy loss NIST ESTAR (water); the track is a multiple-scattering simulation whose mean net distance was compared with published Monte Carlo results (PeneloPET, GATE); e⁺–ν angular correlation Jackson-Treiman-Wyld 1957; mean ranges from the book; 125 ps / 142 ns, 0.5° FWHM from the book; three-photon fraction Mercurio 2006; ortho-positronium lifetime in water Mercolli 2025. Deflection over the first 20% of the track agrees with the Highland formula (PDG) within ±12%; the drawn kinked track uses the energy-loss (~25 keV) points.
How to use: one click pauses and resumes; in the first step ↑ ↓ or a row on the right change the radionuclide, in the annihilation step ↑ ↓ switch the ring diameter (80 / 30 cm); ← → change steps.
Positron, positronium and annihilation · Open in Nuclear Medicine Explained ↗