An ITLC-SG strip is developed first in acetone, then in saline: as the solvent rises, free pertechnetate (and in saline the Tc complex) moves to the front while the rest stays at the origin; the strip is cut and the halves counted. In the last step radiochemical purity is calculated from both results and compared with the 90% limit; three preparation scenarios can be chosen.
Step 1/3
Acetone strip
A drop of the preparation is spotted on an ITLC-SG strip, which is placed in acetone. As the solvent rises, free pertechnetate moves with it to the front (Rf 1); the Tc complex and colloidal Tc stay at the origin (Rf 0). The strip is cut in the middle and both halves counted: the fraction at the front is free TcO₄⁻. Change the preparation.
Key point: acetone measures free pertechnetate (moves to the front), saline measures colloidal Tc (stays at the origin); RCP = 100 − free TcO₄⁻ − colloidal Tc, and for most kits it must be at least 90%.
Source: book (ITLC-SG + acetone: TcO₄⁻ Rf 1, complex and colloid Rf 0; ITLC-SG + saline or 85% methanol: TcO₄⁻ and complex Rf 1, colloid Rf 0; RCP = 100 − free − colloid; limit ≥90%; example 4% / 2% → 94%). Percentages of the other two scenarios illustrative. Lipophilic complexes (e.g. MIBI, HMPAO) need different solvent systems.
How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the preparation scenario; ← → change steps.
Radiochemical purity by ITLC · Open in Nuclear Medicine Explained ↗