Basic sciences · 2 steps

Coincidence timing window and count rates

True coincidences rise linearly with activity and level off once the window covers the arrival difference; randoms rise linearly with window width and with the square of activity, so noise grows at high activity and with a wide window.

Photons reaching two opposed detectors flow along a timeline; if a photon falls in the other detector within the window around a photon, a coincidence is recorded: true if from the same annihilation, random if from different ones. Rates by window width and activity are plotted.

Continuous stream · no seeking

Key point: true coincidences rise linearly with activity and level off once the window covers the arrival difference; randoms rise linearly with window width and with the square of activity, so noise grows at high activity and with a wide window.

Source: book (T ∝ A, R = 2τ·S₁·S₂ ∝ A², window 6–12 ns conventional, ~3–6 ns TOF; scatter 30–40% in 3D). The window width is shown as 2τ (Cherry et al.). Model: photons of one annihilation arrive at most ±2 ns apart (~60 cm patient), timing uncertainty σ 0.25 ns; singles Poisson. On the timeline the time between events is shortened for display (in reality ~10⁵–10⁶ singles/s per detector); curves relative.

How to use: a single click on the drawing pauses and resumes; ↑ ↓ change the window width in the first step and the activity in the second; ← → change steps.

Did this animation help you understand the topic?

Presentation mode ↗The animation alone, sized for full screen; for projecting in lectures and talks. ← → step, space plays/pauses.

Embed

Paste this code into your own page (course site, slides, learning platform). The animation arrives working, with its source written underneath.

Stage

In the book

In the book this animation runs in place of these figures: