Quality, innovations and statistics · 3 steps

Gamma camera quality control

Daily energy peak and uniformity checks and the weekly bar phantom ensure the camera gives accurate, uniform and linear images.

Left: the ±10% window on the energy spectrum and a Tc-99m photopeak sliding with gain drift; a flood image building up as counts accumulate (normal or with a PMT failure); a four-quadrant bar phantom. Right: the share of counts in the window, UFOV and CFOV integral uniformity, and which quadrant is resolved.

Continuous stream · no seeking

Daily energy peak and uniformity checks and the weekly bar phantom ensure the camera gives accurate, uniform and linear images.

Source: test methods and acceptance values from the book (27.1, EANM); uniformity calculated as in NEMA NU 1. Camera size, PMT failure magnitude and counts representative.

How to use: a single click on the drawing pauses and resumes; in the uniformity and bar phantom figures ↑ ↓ or the buttons below switch the state (normal / fault); use the time bar below to go to any moment.