Basic sciences · 6 steps
Positron, annihilation and coincidence
When two 511 keV photons are detected within the timing window the event is placed on the line joining the two detectors; scatter, randoms and prompt gamma coincidences create false lines, and TOF narrows the position along the line.
In a PET ring: positron range, annihilation photons, coincidence and line of response (LOR), scattered, random, multiple and prompt gamma events, and image formation (TOF).
Key point: when two 511 keV photons are detected within the timing window the event is placed on the line joining the two detectors; scatter, randoms and prompt gamma coincidences create false lines, and TOF narrows the position along the line.
Source: book (mean positron ranges, 180° ± 0.25° deviation, T ∝ A, R ∝ τ·A²); prompt gamma: Conti and Eriksson, EJNMMI Phys 2016; ENSDF (I-124 603 keV). Ring 80 cm, to scale; event-type proportions schematic.
How to use: a single click on the drawing pauses and resumes; in the range step ↑ ↓ or the rows of the list pick the radionuclide; in the image steps drag the lesion to move it, and with TOF on ↑ ↓ change the timing resolution; ← → change steps.
Presentation mode ↗The animation alone, sized for full screen; for projecting in lectures and talks. ← → step, space plays/pauses.