Basic sciences · 3 steps

SPECT artefacts

SPECT assumes all projections are consistent; patient motion gives a break and double contour, a fixed detector defect a ring, and a centre-of-rotation shift a blurred or ring-shaped point source.

Three SPECT artefacts are shown with real projection and filtered back-projection: a sinogram break and double contour with patient motion halfway through acquisition, a sinogram stripe and a ring in the slice with detector non-uniformity, and blurring of a point source that turns into a ring with a centre-of-rotation (COR) error. The size of the error is chosen in each step.

Continuous stream · no seeking

Key point: SPECT assumes all projections are consistent; patient motion gives a break and double contour, a fixed detector defect a ring, and a centre-of-rotation shift a blurred or ring-shaped point source.

Source: book (sinogram break, double contour/streaks; non-uniformity → ring; COR: blurred if small, doughnut if large; COR test every 3-6 months). Calculation: same projector as the tomography animation, 64 × 64 (4 mm pixels), 96 angles / 360°, ramp-filtered FBP with 3 × 3 smoothing, noise-free. Shift, defect and COR values illustrative.

How to use: ↑ ↓ or the buttons below change the size of the fault (shift, sensitivity loss, centre-of-rotation error); the sinogram and the slice are recomputed for each choice; ← → change steps.