Radionuclide therapy · 4 steps

TARE dosimetry: MAA, shunt, partition model, DVH

MAA predicts the sphere distribution but is not a copy; planar measurement overestimates the shunt; the partition model splits activity by T/N and mass; a DVH shows the cold volume that the mean hides.

Also in日本語한국어中文

TARE dosimetry in four steps. First, in the same liver, Tc-99m MAA and then Y-90 glass spheres flow from the right hepatic artery; when the Y-90 tip is advanced to the anterior branch, the posterior sector receives no spheres and the regional shares diverge. Second, lung and liver regions are drawn on anterior and posterior planar images and the spill of liver counts into the lung is shown; mean planar, SPECT/CT and realised shunts are compared and the case's lung dose is shown against the 30 and 50 Gy limits. Third, activity is split between lungs, tumour and non-tumoural liver; as T/N changes the two doses move in opposite directions. Fourth, the voxel doses of a tumour slice and the cumulative dose-volume histogram are drawn together as the threshold moves, and D70 and V100 are read.

Continuous stream · no seeking

Key idea: MAA predicts the sphere distribution but is not a copy; planar measurement overestimates the shunt; the partition model splits activity by T/N and mass; a DVH shows the cold volume that the mean hides.

Source: EANM 2022 procedure guideline (49.67 Gy·kg/GBq, lung 30/50 Gy); EANM dosimetry SOP (Chiesa et al. 2021); Wondergem et al. 2013; Keane et al. 2026 (DOSISPHERE-01 ancillary study); Lopez et al. 2022; Young et al. 2023; Pianka et al. 2024. Case is the TARE tool's default (glass, tumour 150 mL, non-tumoural 850 mL, T/N 4, shunt 6%, 1.64 GBq). Particle and voxel counts representative, not to scale.

How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the Y-90 catheter tip, the measurement method or the T/N ratio; ← → change the step.

Speed
Test yourself:Chapter 26 · question 23
Did this animation help you understand the topic?

Presentation mode ↗The animation alone, sized for full screen; for projecting in lectures and talks. ← → step, space plays/pauses.

Embed

Paste this code into your own page (course site, slides, learning platform). The animation arrives working, with its source written underneath.

Stage

In the book

In the book this animation runs in place of these figures:

  • Figure 26.12a · Chapter 26

Author: Uğurcan Doğan, MD →

Step 1/4

MAA versus microspheres

MAA is given before Y-90 from the same catheter position and predicts where the spheres will go. With the same tip the two distributions match; if the tip is advanced, or tumour load is high, the prediction misses at segment level.

MAA, then Y-90: the same liverMAA flowing from the right hepatic arteryMAA (Tc-99m) · 10-90 µm, irregularY-90 glass · 20-30 µmtumour 150 mLY-90 tip: same position as MAA
Share of particles by region (%)TumourAnterior sector (V, VIII)Posterior sector (VI, VII)Upper bar MAA, lower bar Y-90.Same tip: shares match, the prediction holds.Advanced tip: the posterior sector gets no spheres.Both procedures need the same catheter tip.