Basic sciences · 4 steps

Tomographic reconstruction

Simple backprojection gives a star artefact and blurring; the ramp filter corrects this but raises noise; iterative methods fit the image step by step to the measured projections.

Projections and sinogram of a phantom and images reconstructed with simple backprojection, filtered backprojection (FBP) and MLEM/OSEM, computed in the browser.

Step 1/4

Projection and sinogram

The detector rotates around the patient; at each angle the sum of all activity along that direction (line integral) gives a projection profile. Stacking the profiles in angle order gives the sinogram. A point source at the centre is seen in the same place at every angle and draws a straight vertical line; an off-centre point draws a sine curve, hence the name.

Object:

The calculations are real: 64 × 64 matrix, projections over 360° (parallel beam), ramp (Ram-Lak) filter, MLEM/OSEM. The phantom has a body, two hot lesions and one cold region. Noisy data have Poisson noise added; total counts are schematic.

Continuous stream · no seeking

Key point: simple backprojection gives a star artefact and blurring; the ramp filter corrects this but raises noise; iterative methods fit the image step by step to the measured projections.

Source: book; 64 × 64 matrix, 360° parallel-beam projections, Ram-Lak filter, MLEM/OSEM (evenly spaced subsets), Poisson noise; total counts schematic.

How to use: a single click on the images pauses and resumes; ↑ ↓ or the buttons above change that step's choice (object, number of projections, number of subsets); ← → change steps.

Did this animation help you understand the topic?

Presentation mode ↗The animation alone, sized for full screen; for projecting in lectures and talks. ← → step, space plays/pauses.

Embed

Paste this code into your own page (course site, slides, learning platform). The animation arrives working, with its source written underneath.

Stage

In the book

In the book this animation runs in place of these figures: