Total-body PET

The ~40-fold sensitivity of total-body PET comes both from seeing the whole body at once and from catching oblique photon pairs, and it can be spent on lower dose, shorter time or better quality.

Left: annihilations in a patient seen from the side and photon pairs caught or lost depending on the axial FOV, then conventional and total-body PET images building up. Right: total-body and single-organ geometric efficiency of four systems relative to conventional, and how sensitivity, activity, time and decay set the counts.

Continuous stream · no seeking

The ~40-fold sensitivity of total-body PET comes both from seeing the whole body at once and from catching oblique photon pairs, and it can be spent on lower dose, shorter time or better quality.

Source: FOV lengths, ~40-fold gain and ways of use from the book (28.4). Geometric efficiency computed on this page by Monte Carlo (no attenuation, scatter or acceptance-angle limit); phantom and counts representative; F-18 T½ 109.8 min.

How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the system's axial field of view in the first step and the way the sensitivity is used in the second; ← → change steps.