ITLCによる放射化学的純度(radiochemical purity)
アセトンは遊離の過テクネチウム酸(先端へ移動)を、生理食塩水はコロイド状Tc(原点に残る)を測る。RCP = 100 − 遊離TcO₄⁻ − コロイド状Tcであり、ほとんどのキットでは90%以上が必要である。
タイトルと要約は機械支援による翻訳(下書き)です。アニメーション本体は英語です。 英語の完全版ページ →
An ITLC-SG strip is developed first in acetone, then in saline: as the solvent rises, free pertechnetate (and in saline the Tc complex) moves to the front while the rest stays at the origin; the strip is cut and the halves counted. In the last step radiochemical purity is calculated from both results and compared with the 90% limit; three preparation scenarios can be chosen.
Step 1/5
Preparation and running
The tank is filled a few mm deep with the solvent from the product information and closed so the vapour saturates. Origin (1-2 cm from the bottom) and front lines are drawn in pencil and one small drop is placed on the origin. The strip goes in with the spot above the solvent; when the solvent reaches the front it is removed and dried.
Key point: acetone measures free pertechnetate (moves to the front), saline measures colloidal Tc (stays at the origin); RCP = 100 − free TcO₄⁻ − colloidal Tc, and for most kits it must be at least 90%.
Source: book (ITLC-SG + acetone: TcO₄⁻ Rf 1, complex and colloid Rf 0; ITLC-SG + saline or 85% methanol: TcO₄⁻ and complex Rf 1, colloid Rf 0; RCP = 100 − free − colloid; limit ≥90%; example 4% / 2% → 94%). Percentages of the other two scenarios illustrative. Lipophilic complexes (e.g. MIBI, HMPAO) need different solvent systems.
How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the preparation scenario (the error mode in the errors step); ← → change steps.