SPECT収集と回転軌道
対向する二つの検出器は180°の回転で360°分のデータを収集する。心臓検査ではL字配置(L-mode)で心臓に近い前方180°のみを収集する。体輪郭軌道(body-contour orbit)は検出器を患者に近づけ、分解能を保つ。
タイトルと要約は機械支援による翻訳(下書き)です。アニメーション本体は英語です。 英語の完全版ページ →
Gamma camera heads rotating around a thoracic cross-section: 180° opposed and L-mode configurations, the angles acquired, and system resolution on circular and body-contour orbits are shown.
Step 1/3
180° opposed
Two detectors face each other on opposite sides of the patient and rotate in steps; one projection is acquired at each stop. One covers the front half, the other the back half: a 180° rotation collects 360° of data. The same data take twice as long with one head. Used for bone, brain and whole-body SPECT.
Key point: two opposed heads collect 360° of data in a 180° rotation; in cardiac imaging L-mode covers only the anterior 180° near the heart; a body-contour orbit keeps the detector close to the patient and preserves resolution.
Source: book (dual head 180°/90°, cardiac RAO 45° → LPO 45°, 60–128 projections over 360° and 32–64 for cardiac, 15–40 s per projection). Example: 128 and 64 projections, 20 s per projection, rotation time excluded. Body section 36 × 24 cm (illustrative), collimator 1.5 cm from the body. Resolution with the LEHR values of the collimator animation: Rc = d·(L + z)/L, Rs = √(Ri² + Rc²) (d 1.5 mm, L 35 mm, Ri 3.5 mm). Drawing to scale; heads with 40 cm field of view.
How to use: a single click on the drawing pauses and resumes; in the orbit step ↑ ↓ switch between the circular and the body-contour orbit; ← → change steps.