MIRD: 유효 반감기와 누적 방사능(cumulated activity)
유효 반감기는 물리적 반감기와 생물학적 반감기보다 모두 짧다(1/Teff = 1/Tphys + 1/Tbiol). 누적 방사능은 곡선하 면적이며, Ã = 1.443 × A₀ × Teff이다.
제목과 요약은 기계 보조 번역(초안)입니다. 애니메이션은 영어로 되어 있습니다. 영어 전체 페이지 →
Atoms in an organ are lost by physical decay or biological excretion; physical, biological and effective retention curves and the area under the activity–time curve (cumulated activity) are shown.
Key point: the effective half-life is shorter than both the physical and biological half-lives (1/Teff = 1/Tphys + 1/Tbiol); cumulated activity is the area under the curve, Ã = 1.443 × A₀ × Teff.
Source: book (example Tphys 6 h, Tbiol 24 h → Teff 4.8 h); A₀ = 100 MBq and f = 1 schematic; Tc-99m 6.01 h, I-123 13.2 h, I-131 8.02 d.
How to use: one click on the organ scene pauses and resumes; move over or drag along the graph; in the free step ↑ ↓ change the biological half-life smoothly; ← → change steps.