部分容积效应

病灶小于2-3倍FWHM时,热病灶被低估(溢出,spill-out),冷病灶被高估(溢入,spill-in);这种损失用恢复系数(recovery coefficient)描述,并可校正。

标题和摘要为机器辅助翻译(草稿);动画内容为英文。 英文完整页面 →

A section through the centre of a uniform sphere, blurred by the system FWHM, is shown side by side for a hot (4:1) and a cold (0) lesion; on the right the profiles are compared with the true values. The second step shows the recovery coefficient (maximum and mean) against lesion diameter / FWHM and where the chosen lesion lies.

Continuous stream · no seeking

Key point: when the lesion is smaller than 2-3 × FWHM a hot lesion is underestimated (spill-out) and a cold lesion overestimated (spill-in); the loss is described by the recovery coefficient and can be corrected.

Source: book (PVE < 2-3 × FWHM, spill-out / spill-in, RC = measured/true, corrected = measured/RC). Calculation: convolution of a uniform sphere with a 3D Gaussian (analytic formula; σ = FWHM/2.355); hot lesion 4:1, cold 0:1; mean RC over the sphere volume. No noise or voxel size effects.

How to use: ↑ ↓ or the buttons below change the lesion diameter; the second row picks the resolution (FWHM); ← → change steps.