In a gamma camera the scintillation light spreads to the PMTs, position is found by a weighted mean and energy by the summed signal; the fraction of photons absorbed and the light spread by crystal thickness are shown.
Key point: the weighted mean of the PMT signals gives the position and their sum the energy; a thick crystal catches more photons but spreads the light wider, so the position becomes less certain.
Source: book (NaI(Tl) 3/8 inch, 3.67 g/cm³; X = Σ(Xᵢ·Sᵢ)/ΣSᵢ, Z = ΣSᵢ; Tc-99m window 126–154 keV). Absorption from the NIST XCOM NaI total attenuation coefficient (coherent included); the book's ~30% at 364 keV is the photopeak efficiency, as part of the interacting photons escape by Compton scattering. In the crystal-thickness step the side view is drawn at 2× vertical scale (light guide 12 mm). Light spot: point light source, reflections ignored, FWHM ≈ 1.53 × distance to the PMT plane. Model assumptions: 37 PMTs, 76 mm pitch, 12 mm light guide, ~930 photoelectrons at 140 keV, Poisson noise; PMTs below 10% of the largest signal ignored (like the threshold in modern cameras); scattered photons ~25% of events, 95–125 keV (illustrative).
How to use: a single click on the drawing pauses and resumes; in the crystal thickness step ↑ ↓ change the thickness, the buttons below change thickness and energy; ← → change steps.
Anger logic and crystal thickness · Open in Nuclear Medicine Explained ↗