In the ECG strip each R-R interval is divided into frames, a short-axis section of the left ventricle beats, and LVEF is calculated from the frame-averaged volume curve. The second step rejects beats outside the mean R-R ±20% in a variable rhythm and shows the R-R distribution. The third step bins an irregular breathing signal by phase or amplitude and compares the diaphragm range and time share of each bin.

Continuous stream · no seeking

Key point: gating freezes motion by sorting counts by cardiac or respiratory phase; with irregular R-R, beat rejection costs counts, and with phase-based respiratory binning irregular breaths blur the image; end-expiration is the most stable phase.

Source: book (R-R in 8 or 16 frames; LVEF = (EDV − ESV)/EDV; EDV usually frame 1, ESV 3-4 of 8; acceptance window mean R-R ±20%; breathing 4-10 phases, phase- and amplitude-based, end-expiration most stable, duty cycle). Volume curve illustrative (EDV 120, ESV 50 mL, minimum at 37% of R-R); myocardial volume kept constant in the short-axis ring. R-R sequences illustrative (PVC 620/1380 ms; AF 500-1150 ms). Breathing: Lujan 1999 model d = A·cos⁴(πt/τ), τ = 4 s, amplitudes 1.4-3.0 cm illustrative.

How to use: a single click on the drawing pauses and resumes; the buttons below the drawing pick the number of frames, the rhythm and the method; ← → change steps and selections.

Did this animation help you understand the topic?

ECG and respiratory gating · Open in Nuclear Medicine Explained ↗