Basic sciences · 3 steps
ECG and respiratory gating
Gating freezes motion by sorting counts by cardiac or respiratory phase; with irregular R-R, beat rejection costs counts, and with phase-based respiratory binning irregular breaths blur the image; end-expiration is the most stable phase.
In the ECG strip each R-R interval is divided into frames, a short-axis section of the left ventricle beats, and LVEF is calculated from the frame-averaged volume curve. The second step rejects beats outside the mean R-R ±20% in a variable rhythm and shows the R-R distribution. The third step bins an irregular breathing signal by phase or amplitude and compares the diaphragm range and time share of each bin.
Key point: gating freezes motion by sorting counts by cardiac or respiratory phase; with irregular R-R, beat rejection costs counts, and with phase-based respiratory binning irregular breaths blur the image; end-expiration is the most stable phase.
Source: book (R-R in 8 or 16 frames; LVEF = (EDV − ESV)/EDV; EDV usually frame 1, ESV 3-4 of 8; acceptance window mean R-R ±20%; breathing 4-10 phases, phase- and amplitude-based, end-expiration most stable, duty cycle). Volume curve illustrative (EDV 120, ESV 50 mL, minimum at 37% of R-R); myocardial volume kept constant in the short-axis ring. R-R sequences illustrative (PVC 620/1380 ms; AF 500-1150 ms). Breathing: Lujan 1999 model d = A·cos⁴(πt/τ), τ = 4 s, amplitudes 1.4-3.0 cm illustrative.
How to use: a single click on the drawing pauses and resumes; the buttons below the drawing pick the number of frames, the rhythm and the method; ← → change steps and selections.
Presentation mode ↗The animation alone, sized for full screen; for projecting in lectures and talks. ← → step, space plays/pauses.
Embed
Paste this code into your own page (course site, slides, learning platform). The animation arrives working, with its source written underneath.
In the book
In the book this animation runs in place of these figures: