The energy spectrum of photons reaching the detector builds up event by event (photopeak, Compton, backscatter, lead X-ray, escape and sum peaks); FWHM measurement and the pulse-height analyser window are shown.
Key point: the photopeak is full energy absorption; Compton, backscatter, escape and X-ray peaks are partial or secondary events. The energy window rejects photons scattered in the patient; a narrower window means less scatter but also fewer counts.
Source: Compton edge and backscatter from formula (Tc-99m 49.8 / 90.7, Cs-137 477.6 / 184.3 keV); Pb Kα₁ 74.97, Kβ₁ 84.94 keV; I K X-ray 28.6 keV; Compton angles from the Klein–Nishina distribution. NaI FWHM 10% at 140 keV (∝ 1/√E), CZT 5.5%. Event-type fractions are representative. HPGe FWHM: 0.5–1.0 keV at 122 keV (Mirion and ORTEC data sheets, 2025); 0.84 keV here. Escape peaks follow the detector's own K X-ray: I 28.6; Cd 23.2 / Te 27.5; Ge 9.9 keV (Bearden 1967).
How to use: a single click on the drawing pauses and resumes; the buttons below the drawing pick the source, the detector or the window width; ← → change steps and selections.
Energy spectrum, resolution and the PHA window · Open in Nuclear Medicine Explained ↗