Basic sciences · 3 steps
Energy spectrum, resolution and the PHA window
The photopeak is full energy absorption; Compton, backscatter, escape and X-ray peaks are partial or secondary events. The energy window rejects photons scattered in the patient; a narrower window means less scatter but also fewer counts.
The energy spectrum of photons reaching the detector builds up event by event (photopeak, Compton, backscatter, lead X-ray, escape and sum peaks); FWHM measurement and the pulse-height analyser window are shown.
Key point: the photopeak is full energy absorption; Compton, backscatter, escape and X-ray peaks are partial or secondary events. The energy window rejects photons scattered in the patient; a narrower window means less scatter but also fewer counts.
Source: Compton edge and backscatter from formula (Tc-99m 49.8 / 90.7, Cs-137 477.6 / 184.3 keV); Pb Kα₁ 74.97, Kβ₁ 84.94 keV; I K X-ray 28.6 keV; Compton angles from the Klein–Nishina distribution. NaI FWHM 10% at 140 keV (∝ 1/√E), CZT 5.5%. Event-type fractions are representative. HPGe FWHM: 0.5–1.0 keV at 122 keV (Mirion and ORTEC data sheets, 2025); 0.84 keV here. Escape peaks follow the detector's own K X-ray: I 28.6; Cd 23.2 / Te 27.5; Ge 9.9 keV (Bearden 1967).
How to use: a single click on the drawing pauses and resumes; the buttons below the drawing pick the source, the detector or the window width; ← → change steps and selections.
Presentation mode ↗The animation alone, sized for full screen; for projecting in lectures and talks. ← → step, space plays/pauses.
Embed
Paste this code into your own page (course site, slides, learning platform). The animation arrives working, with its source written underneath.
In the book
In the book this animation runs in place of these figures: