Left: a section of the gamma probe (crystal, shield, light detector, readout), scanning over the skin, and ex vivo node counting. Right: detector comparison, count profiles by position for narrow and wide collimation, and node counts with the 10% threshold.

Step 1/3

Probe structure

A gamma probe is a hand-held detector with a small crystal at its tip. A tungsten or lead shield around the crystal stops radiation from the side, so only photons from where the probe points are counted. The count rate is shown in cps and as an audio tone proportional to the counts. In scintillation crystals (NaI, CsI) a light detector turns light into an electrical signal; CZT is a semiconductor that converts the photon directly into charge.

Gamma probe structure (section)1150 cpsSLN photonsside photon stopped by the shieldcrystal: NaI(Tl)collimation + side shield(tungsten/lead)light detector +electronicsdigital readout + audio tone
Detector choiceDetectorAdvantageDisadvantageNaI(Tl)High sensitivity,widely usedBulky,hygroscopicCsI(Tl)Compact,robustModerate energyresolutionCZTExcellent energyresolution, compact,room temperatureHigh costScintillation probes are more sensitive and better athigh energies (I-131, 511 keV); semiconductor probesare smaller with better energy resolution.
Continuous stream · no seeking

Narrow collimation and side shielding give the probe its directional sensitivity; every node with ≥10% of the highest ex vivo count is an SLN.

Source: probe structure, detector comparison and the 10% rule from book 22.5; ex vivo counts are the book's example values. Scanning curves from a representative angular response model.

How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the detector crystal in the first step and the acceptance angle in the second; ← → change steps.