Clinical applications · 3 steps
Gamma probe and the 10% rule
Narrow collimation and side shielding give the probe its directional sensitivity; every node with ≥10% of the highest ex vivo count is an SLN.
Left: a section of the gamma probe (crystal, shield, light detector, readout), scanning over the skin, and ex vivo node counting. Right: detector comparison, count profiles by position for narrow and wide collimation, and node counts with the 10% threshold.
Narrow collimation and side shielding give the probe its directional sensitivity; every node with ≥10% of the highest ex vivo count is an SLN.
Source: probe structure, detector comparison and the 10% rule from book 22.5; ex vivo counts are the book's example values. Scanning curves from a representative angular response model.
How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the detector crystal in the first step and the acceptance angle in the second; ← → change steps.
Presentation mode ↗The animation alone, sized for full screen; for projecting in lectures and talks. ← → step, space plays/pauses.