Formation and collection of ion pairs in a gas-filled detector, the six operating regions by voltage, and the discharge and dead time of a Geiger-Müller counter are shown.
Step 1/3
Ion pairs and collection
A charged particle ionizes the gas as it passes; in argon one ion pair is formed per 26.4 eV. In the electric field electrons drift to the thin central anode wire and positive ions to the cylindrical cathode wall. Electrons are collected in µs, ions about 1000 times more slowly. A 5.49 MeV α that stops completely leaves about 2×10⁵ pairs; a 1 MeV β crossing the chamber only ~5×10² (6 cm chamber).
Pick the particle below the drawing or with ↑ ↓; in step 2 you can drag on the chart to set the voltage. The small window on the left shows the region next to the wire at micro scale.
Key point: an ion chamber collects the charge as formed, a proportional counter multiplies it but keeps energy information, and a GM counter gives the same large pulse for every event and cannot discriminate energy.
Source: argon W = 26.4 eV per pair (ICRU); α range in argon 4.5 cm (NIST ASTAR, 1 atm), chamber cross-section 6 cm; β energy loss NIST ESTAR (1 MeV, 2.29 keV/cm); gas multiplication and GM dead time from the book. Time scales slowed and not to scale. Near the wire: Diethorn relation, for P-10 gas (90% Ar + 10% CH₄) K = 4.8×10⁴ V/cm·atm, ΔV = 23.6 V (Wolff 1974; Knoll); example wire radius 25 µm, cathode radius 3 cm.