Basic sciences · 3 steps

Gas-filled detectors

An ion chamber collects the charge as formed, a proportional counter multiplies it but keeps energy information, and a GM counter gives the same large pulse for every event and cannot discriminate energy.

Formation and collection of ion pairs in a gas-filled detector, the six operating regions by voltage, and the discharge and dead time of a Geiger-Müller counter are shown.

Continuous stream · no seeking

Key point: an ion chamber collects the charge as formed, a proportional counter multiplies it but keeps energy information, and a GM counter gives the same large pulse for every event and cannot discriminate energy.

Source: argon W = 26.4 eV per pair (ICRU); α range in argon 4.5 cm (NIST ASTAR, 1 atm), chamber cross-section 6 cm; β energy loss NIST ESTAR (1 MeV, 2.29 keV/cm); gas multiplication and GM dead time from the book. Time scales slowed and not to scale. Near the wire: Diethorn relation, for P-10 gas (90% Ar + 10% CH₄) K = 4.8×10⁴ V/cm·atm, ΔV = 23.6 V (Wolff 1974; Knoll); example wire radius 25 µm, cathode radius 3 cm.

How to use: a single click on the drawing pauses and resumes; ↑ ↓ switch the particle (α / β); in the voltage step drag on the right-hand chart to set the voltage; ← → change steps.