In a coronal section the diaphragm moves with breathing; PET sees the average of minutes of breathing, CT a single moment. The moment of the CT is chosen; the second step shows how the difference causes a cold band at the liver dome or a hot artefact at the lung base in CT attenuation correction.
Step 1/2
PET average, CT moment
The diaphragm moves up and down with each breath, staying longer at end-expiration. PET collects a bed position over minutes and sees the average diaphragm position. CT is taken in seconds and catches a single moment of breathing. Choose when the CT was taken: the gap between the two lines is the misregistration.
Key point: CT is a snapshot, PET a breathing average; a CT in deep inspiration gives a cold band at the liver dome, a diaphragm higher than in PET a hot artefact at the lung base; CT in shallow breathing or gating prevents this.
Source: book (CT instantaneous, PET average; cold band at the liver dome or hot artefact at the lung base; solution CT in shallow breathing, respiratory gating; MRAC µ values lung 0.0224, soft tissue 0.100 cm⁻¹). Breathing curve from the Lujan et al. 1999 model (d = A·cos⁴(πt/τ), more time at expiration; mean 3/8·A). Amplitude 2 cm, period 4 s and CT positions (4/2/0 cm, shallow ≈ mean) illustrative.
How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the breathing moment of the CT; ← → change steps.