Respiratory misregistration

CT is a snapshot, PET a breathing average; a CT in deep inspiration gives a cold band at the liver dome, a diaphragm higher than in PET a hot artefact at the lung base; CT in shallow breathing or gating prevents this.

In a coronal section the diaphragm moves with breathing; PET sees the average of minutes of breathing, CT a single moment. The moment of the CT is chosen; the second step shows how the difference causes a cold band at the liver dome or a hot artefact at the lung base in CT attenuation correction.

Continuous stream · no seeking

Key point: CT is a snapshot, PET a breathing average; a CT in deep inspiration gives a cold band at the liver dome, a diaphragm higher than in PET a hot artefact at the lung base; CT in shallow breathing or gating prevents this.

Source: book (CT instantaneous, PET average; cold band at the liver dome or hot artefact at the lung base; solution CT in shallow breathing, respiratory gating; MRAC µ values lung 0.0224, soft tissue 0.100 cm⁻¹). Breathing curve from the Lujan et al. 1999 model (d = A·cos⁴(πt/τ), more time at expiration; mean 3/8·A). Amplitude 2 cm, period 4 s and CT positions (4/2/0 cm, shallow ≈ mean) illustrative.

How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the breathing moment of the CT; ← → change steps.

Did this animation help you understand the topic?

Presentation mode ↗The animation alone, sized for full screen; for projecting in lectures and talks. ← → step, space plays/pauses.

Embed

Paste this code into your own page (course site, slides, learning platform). The animation arrives working, with its source written underneath.

Stage

In the book

In the book this animation runs in place of these figures: