A 140 keV γ photon converted to light in a NaI(Tl) crystal, to electrons at the photocathode and multiplied at the dynodes; carrier numbers and photopeak width in NaI and CZT are shown.
Key point: in a scintillation detector the energy passes through several conversions and carriers are lost at each; at 140 keV ~900 photoelectrons give ~10% resolution, CZT's ~30,000 carriers ~5–6%.
Source: book (NaI ~38 photons/keV, ~415 nm, ~230 ns; quantum efficiency 20–30%; dynode ×3–6, 10–12 dynodes; CZT 3–5 eV per carrier). Assumptions: ~70% of light reaches the photocathode, QE 25%, ×4; CZT W ≈ 4.6 eV (published value). In CZT other effects such as charge trapping also limit resolution. Micro scale: I K shell 33.17 keV, I Kα 28.6 keV (Bearden 1967); photoelectron path from the integral of the NIST ESTAR stopping power (NaI); mean free path at 28.6 keV in NaI 0.33 mm (NIST XCOM). Magnetic field: r = m·v/(e·B), ≈ 11 µm for a 100 eV electron at 3 T (~100 V between dynodes assumed); orbits are drawn enlarged.
How to use: a single click on the drawing pauses and resumes; ← → change steps; in the photocathode and dynodes step ↑ ↓ or the buttons below switch the MR field on and off.
Scintillation detector and PMT · Open in Nuclear Medicine Explained ↗