Collimator types and the effect of distance

The collimator sets the image geometry; with a parallel-hole collimator resolution worsens with distance while sensitivity stays the same; a pinhole inverts the image and magnifies nearby objects.

Photons from an object are selected by the collimator and those passing form the image in the crystal; parallel-hole, pinhole, converging and diverging collimators and the effect of source–collimator distance are shown.

Continuous stream · no seeking

Key point: the collimator sets the image geometry; with a parallel-hole collimator resolution worsens with distance while sensitivity stays the same; a pinhole inverts the image and magnifies nearby objects.

Source: book; LEHR d = 1.5 mm, L = 35 mm (manufacturer value), Ri = 3.5 mm; Rc = d·(L + z)/L, Rs = √(Ri² + Rc²); pinhole M = f/z, f = 20 cm example value. Photon numbers and angles not to scale. Septa step: LEHR 1.5 / 0.2 / 35 mm and MEGP 3.0 / 1.05 / 58 mm (GE Discovery; Wright et al., Phys Med Biol 2017), HE 4.0 / 2.0 / 59.7 mm (Siemens Symbia); lead attenuation coefficients from NIST XCOM. This is a 2D cross-section simulation; true (3D) penetration fractions differ (Monte Carlo: LEHR + I-131 ~50%, HEGP + I-131 ~18%; Sawant et al., J Nucl Med Technol 2025).

How to use: a single click on the drawing pauses and resumes; in the distance steps drag the object or use ← → to move it closer or farther; in the septa step ↑ ↓ change the radionuclide.