Quality control · 3 of 8

Daily dose calibrator check

Awaiting physician review

To show, before the first measurement of the day, that the calibrator works accurately and stably, so that the activity given to the patient is measured correctly.

What you need

  • Long-lived, certified check source (e.g. Cs-137; Co-57 or Ba-133 are also used)
  • Tongs, source container
  • The calibrator's source holder (dipper/liner)
  • Record form or electronic log

Steps

  1. Physical inspection: check the calibrator, source holder and accessories for damage and contamination.
  2. Check the high voltage as the manufacturer recommends.
  3. Compare the clock with the other clocks in the department (injection room, computers); the time of assay directly affects the activity.
  4. Zero: confirm that the display reads zero with no radioactive source present; record the value before making any adjustment.
  5. Background: with the source holder in place and nearby sources removed, measure the background. If it is higher than usual, look for contamination or a nearby source.
  6. Constancy: place the check source in the holder with tongs and assay it on its own setting (Cs-137 on the Cs-137 setting). It should be within ±5% of the decay-corrected initial value.
  7. Relative response: assay the same source on each radionuclide setting to be used that day (Tc-99m, I-131, F-18…) and compare with previous values.
  8. Record all results; if any is out of tolerance, do not use the calibrator and inform the responsible physicist.

Key points

  • Daily tests: physical inspection, high voltage, clock, zero, background, constancy and relative response. Accuracy, reproducibility and linearity yearly (linearity six-monthly in some programmes); geometry at acceptance and whenever the container type changes.
  • The constancy tolerance in AAPM TG-181 is ±5% of the decay-corrected initial value. Some regulations and older practice use ±10%; the limit in the local programme applies.
  • A drifting zero or rising background usually points to contamination or a fault.

Common mistakes

  • Assaying the check source on the wrong setting
  • Forgetting the decay correction
  • Measuring background with an open source nearby
  • Adjusting the zero without recording it first

Safety. Move the check source with tongs, return it to its container and keep it locked away.

Sources: AAPM TG-181 (2012) · EANM 2017 (Bölüm 9) · IAEA-TECDOC-1856 (2018) · SNMMI/ACNM 5.0 (2025). Last reviewed: 23 November 2026.