Quality control · 2 of 8
Radiochemical purity: instant thin-layer chromatography (ITLC) of Tc-99m kits
Awaiting physician review
To measure the percentage of activity present in the intended chemical form (radiochemical purity, RCP) and separate the free pertechnetate and reduced-hydrolysed Tc-99m (RH-Tc) impurities.
What you need
- The strip (e.g. ITLC-SG) and solvents named in the kit's product information
- Lidded chromatography tank
- Pencil, ruler, scissors
- Small QC syringe and fine needle
- Well counter or TLC scanner
- Record form
Steps
- Fill the tank with the specified solvent to a depth of a few mm, close the lid and let the vapour saturate.
- Mark an origin line in pencil 1-2 cm from the bottom of the strip and a solvent-front line near the top.
- Spot a single small drop on the origin with a fine needle (keep the spot small); dry it if the product information requires.
- Place the strip in the tank with the spot above the solvent surface and close the lid.
- When the solvent reaches the front line, remove the strip and let it dry.
- Cut the strip at the point given in the product information (or read it on a scanner); count each piece for the same time and in the same geometry.
- Calculate the impurity percentage in each system; RCP = 100 − (free pertechnetate %) − (RH-Tc %).
- Compare the result with the kit's acceptance limit and record it; below the limit, the kit is not given to a patient.
Key points
- Most Tc-99m kits need two systems: in one (e.g. acetone or MEK) free pertechnetate moves to the front while the labelled compound and RH-Tc stay at the origin; in the other (e.g. saline) the labelled compound and pertechnetate move and RH-Tc stays at the origin. The strip and solvent to use are given in the kit's product information.
- Most diagnostic kits are expected to reach RCP ≥ 90-95%; a few have a lower limit (e.g. HMPAO > 80%). The limit is in the kit's product information or the Ph. Eur. monograph.
- Low-activity samples are not measured reliably in a dose calibrator; a well counter or TLC scanner is preferred, and the counter must not saturate.
Common mistakes
- Spot immersed in the solvent (washed off)
- Large drop (tailing, poor separation)
- Lid left open (vapour not saturated)
- Wrong strip or solvent
- Counting pieces for different times or in different geometry
Safety. Spot behind a shield; discard strips and syringes as radioactive waste.
Sources: UKRG 2021 · EANM 2019 · IAEA/WHO QAS/24.969 (2025). Last reviewed: 23 November 2026.