Quality control · 2 of 8

Radiochemical purity: instant thin-layer chromatography (ITLC) of Tc-99m kits

Awaiting physician review

To measure the percentage of activity present in the intended chemical form (radiochemical purity, RCP) and separate the free pertechnetate and reduced-hydrolysed Tc-99m (RH-Tc) impurities.

What you need

  • The strip (e.g. ITLC-SG) and solvents named in the kit's product information
  • Lidded chromatography tank
  • Pencil, ruler, scissors
  • Small QC syringe and fine needle
  • Well counter or TLC scanner
  • Record form

Steps

  1. Fill the tank with the specified solvent to a depth of a few mm, close the lid and let the vapour saturate.
  2. Mark an origin line in pencil 1-2 cm from the bottom of the strip and a solvent-front line near the top.
  3. Spot a single small drop on the origin with a fine needle (keep the spot small); dry it if the product information requires.
  4. Place the strip in the tank with the spot above the solvent surface and close the lid.
  5. When the solvent reaches the front line, remove the strip and let it dry.
  6. Cut the strip at the point given in the product information (or read it on a scanner); count each piece for the same time and in the same geometry.
  7. Calculate the impurity percentage in each system; RCP = 100 − (free pertechnetate %) − (RH-Tc %).
  8. Compare the result with the kit's acceptance limit and record it; below the limit, the kit is not given to a patient.

Key points

  • Most Tc-99m kits need two systems: in one (e.g. acetone or MEK) free pertechnetate moves to the front while the labelled compound and RH-Tc stay at the origin; in the other (e.g. saline) the labelled compound and pertechnetate move and RH-Tc stays at the origin. The strip and solvent to use are given in the kit's product information.
  • Most diagnostic kits are expected to reach RCP ≥ 90-95%; a few have a lower limit (e.g. HMPAO > 80%). The limit is in the kit's product information or the Ph. Eur. monograph.
  • Low-activity samples are not measured reliably in a dose calibrator; a well counter or TLC scanner is preferred, and the counter must not saturate.

Common mistakes

  • Spot immersed in the solvent (washed off)
  • Large drop (tailing, poor separation)
  • Lid left open (vapour not saturated)
  • Wrong strip or solvent
  • Counting pieces for different times or in different geometry

Safety. Spot behind a shield; discard strips and syringes as radioactive waste.

Sources: UKRG 2021 · EANM 2019 · IAEA/WHO QAS/24.969 (2025). Last reviewed: 23 November 2026.