Quality control · 1 of 8

Mo-99 breakthrough and Al³⁺ test (every elution)

Awaiting physician review

To show that the eluate contains no Mo-99 (radionuclidic impurity) or aluminium (chemical impurity) above the permitted limit.

What you need

  • Dose calibrator with the manufacturer's Mo-99 assay shield (lead pig)
  • Tongs
  • Aluminium test kit (colour strip and the standard of the pharmacopoeia in use: Ph. Eur. 5 µg/mL, USP 10 µg/mL)
  • Small syringe and needle
  • Record form

Steps

  1. Mo-99: zero the calibrator on the Mo-99 setting. Place the eluate vial in the Mo-99 shield with tongs and assay it on the Mo-99 setting; the lead stops the 140 keV Tc-99m photons, while the 740/780 keV Mo-99 photons pass (the setting corrects for partial absorption).
  2. Remove the vial from the shield and assay it on the Tc-99m setting.
  3. Calculate the ratio: Mo-99 activity / Tc-99m activity. Compare it with the limit of the pharmacopoeia you work under and record it.
  4. Al³⁺: take a small sample of the eluate. Place one drop of eluate at one end of the strip and an equal drop of the aluminium standard (Ph. Eur. 5 µg/mL or USP 10 µg/mL) at the other.
  5. Compare the colour at the centre of the drops: if the eluate spot is paler than the standard, the test passes; if it is the same or darker, the eluate is not used.
  6. Discard the syringe, strip and needle as radioactive waste and record the results in the elution log.

Key points

  • The Mo-99 limit depends on the pharmacopoeia: Ph. Eur. (and the International Pharmacopoeia) 0.1% of total activity (1 kBq Mo-99 per MBq Tc-99m); USP 0.15 kBq/MBq (0.15 µCi/mCi, 0.015%). Both apply at the time of administration.
  • The measurement is made just after elution, but the limit applies at administration: Tc-99m decays faster, so the Mo/Tc ratio rises with time (about 2.3-fold in 8 hours). For the Ph. Eur. limit, an eluate to be used 8 hours later should not exceed about 0.04% at elution.
  • The aluminium limit also differs between pharmacopoeias: Ph. Eur. and the International Pharmacopoeia 5 µg/mL (5 ppm); USP 10 µg/mL. The strip is compared with the standard of the pharmacopoeia in use; the limit in the product information applies. Excess Al³⁺ impairs sulphur colloid and red-cell labelling.
  • Excess Mo-99 adds patient dose without contributing to the image and degrades image quality.

Common mistakes

  • Mixing up the shield or the calibrator setting for the Mo-99 assay
  • Judging the limit at the time of measurement and forgetting the time of administration
  • Using unequal drops in the Al³⁺ test

Safety. Move the vial with tongs; take the test sample behind a shield.

Sources: IAEA-TECDOC-1856 (2018) · IAEA TRS 454 (2006) · USP monografı · Ph. Eur. Al 5 ppm: WHO QAS/17.729 (2017), Ph. Eur. 2891 (BP 2025), TechneLite AU ürün bilgisi · SmPC · EANM 2008, 2019. Last reviewed: 23 November 2026.