Radiopharmacy · 1 of 8

Mo-99/Tc-99m generator elution (dry column, evacuated vial)

Awaiting physician review

To obtain sterile, pyrogen-free sodium pertechnetate (Tc-99m) from the generator while keeping staff dose low.

What you need

  • Sterile 0.9% NaCl (saline) vial, or the generator's internal reservoir
  • Sterile evacuated elution vial (sized to the required volume)
  • Elution vial shield with lead cap
  • Alcohol swab, sterile needle or port cap
  • Gloves, lab coat, body and ring dosimeters
  • Dose calibrator, record form, label

Steps

  1. Put on your dosimeters and gloves. Check the generator's calibration and expiry dates and the time of the last elution.
  2. Wipe the saline vial septum with an alcohol swab and let it dry; seat the vial on the inlet needle (not needed for models with an internal reservoir).
  3. Place the evacuated vial in its shield, wipe the septum with alcohol and let it dry.
  4. Remove the protective cap from the outlet port and push the shielded vial fully onto the outlet needle.
  5. The vacuum draws the saline through the alumina column into the vial. Leave the vial in place for the time the manufacturer specifies (usually a few minutes).
  6. Remove the shielded vial and fit its lead cap; close the outlet port or fit a new sterile needle or cap.
  7. Assay the eluate in the dose calibrator on the Tc-99m setting; record the time, volume and activity and calculate the radioactive concentration (MBq/mL).
  8. Quality control: appearance (clear, colourless), Mo-99 breakthrough and Al³⁺ test (separate card). If the results are acceptable, label the vial: radionuclide, activity, volume, assay time, expiry.

Key points

  • Mo-99 is bound to the alumina (Al₂O₃) column as molybdate (MoO₄²⁻); the pertechnetate formed by decay binds weakly and is washed off with saline.
  • Tc-99m activity builds up again after each elution: about 50% of the maximum at 4.5 hours and 73% at 8 hours, with the maximum at ~23 hours.
  • The eluate's shelf life is given in the product information (8 hours for one manufacturer); when labelling kits, the kit's own limit also applies.
  • The column and fluid path are never removed from their shielding; the whole procedure follows aseptic technique.

Common mistakes

  • Piercing the septum before the alcohol has dried
  • Removing the vial before the vacuum has finished (incomplete elution, low yield)
  • Using a vial that has lost its vacuum
  • Carrying an unshielded vial by hand

Safety. Use the shielded vial and tongs, keep your distance and keep handling time short (the ring dosimeter is on your hand).

Sources: IAEA VirRad · EANM 2019 · SmPC · NCBI 2016 (Medical Isotope Production and Utilization). Last reviewed: 23 November 2026.