Radiopharmacy · 1 of 8
Mo-99/Tc-99m generator elution (dry column, evacuated vial)
Awaiting physician review
To obtain sterile, pyrogen-free sodium pertechnetate (Tc-99m) from the generator while keeping staff dose low.
What you need
- Sterile 0.9% NaCl (saline) vial, or the generator's internal reservoir
- Sterile evacuated elution vial (sized to the required volume)
- Elution vial shield with lead cap
- Alcohol swab, sterile needle or port cap
- Gloves, lab coat, body and ring dosimeters
- Dose calibrator, record form, label
Steps
- Put on your dosimeters and gloves. Check the generator's calibration and expiry dates and the time of the last elution.
- Wipe the saline vial septum with an alcohol swab and let it dry; seat the vial on the inlet needle (not needed for models with an internal reservoir).
- Place the evacuated vial in its shield, wipe the septum with alcohol and let it dry.
- Remove the protective cap from the outlet port and push the shielded vial fully onto the outlet needle.
- The vacuum draws the saline through the alumina column into the vial. Leave the vial in place for the time the manufacturer specifies (usually a few minutes).
- Remove the shielded vial and fit its lead cap; close the outlet port or fit a new sterile needle or cap.
- Assay the eluate in the dose calibrator on the Tc-99m setting; record the time, volume and activity and calculate the radioactive concentration (MBq/mL).
- Quality control: appearance (clear, colourless), Mo-99 breakthrough and Al³⁺ test (separate card). If the results are acceptable, label the vial: radionuclide, activity, volume, assay time, expiry.
Key points
- Mo-99 is bound to the alumina (Al₂O₃) column as molybdate (MoO₄²⁻); the pertechnetate formed by decay binds weakly and is washed off with saline.
- Tc-99m activity builds up again after each elution: about 50% of the maximum at 4.5 hours and 73% at 8 hours, with the maximum at ~23 hours.
- The eluate's shelf life is given in the product information (8 hours for one manufacturer); when labelling kits, the kit's own limit also applies.
- The column and fluid path are never removed from their shielding; the whole procedure follows aseptic technique.
Common mistakes
- Piercing the septum before the alcohol has dried
- Removing the vial before the vacuum has finished (incomplete elution, low yield)
- Using a vial that has lost its vacuum
- Carrying an unshielded vial by hand
Safety. Use the shielded vial and tongs, keep your distance and keep handling time short (the ring dosimeter is on your hand).
Sources: IAEA VirRad · EANM 2019 · SmPC · NCBI 2016 (Medical Isotope Production and Utilization). Last reviewed: 23 November 2026.