Ventilation scintigraphy: Technegas and DTPA aerosol

Awaiting physician review

To deliver an adequate and uniform ventilation agent to the lungs to reach a suitable count rate for imaging without contaminating the room.

What you need

  • Technegas system, argon, graphite crucible, ethanol; or a nebuliser and Tc-99m DTPA
  • Patient administration set (filtered)
  • Oximeter
  • Gamma camera (to monitor the count rate)
  • Gloves, absorbent pad, waste bag

Steps

  1. Teach and rehearse the breathing technique with the patient beforehand; monitor oxygen saturation with an oximeter.
  2. Technegas: turn on the argon, wet the crucible with ethanol and drain it, place it between the contacts (rotate for contact, do not twist). Add 400-900 MBq of pertechnetate in the volume the product and system recommend (standard crucible 0.12-0.17 mL); the meniscus must not overflow.
  3. Run the simmer (~70 °C) and burn (~2750 °C, 15 s) cycle; administer the Technegas within 10 minutes.
  4. With the patient supine, connect the administration set; press the delivery button on inspiration. Stop when the posterior count rate reaches 1500-2500 counts/s; let the patient take a few more breaths through the set so the residual goes to the filter.
  5. DTPA aerosol: load 900-1300 MBq into the nebuliser; the patient breathes through a mouthpiece (nose clipped); about 20-40 MBq reaches the lungs.
  6. Dispose of the set without detaching it from the patient first; put the system through its cleaning and purge cycle.
  7. Start imaging immediately; if ventilation precedes perfusion, set the MAA activity according to the protocol.
  8. Survey surfaces and hands; clean and record any contamination.

Key points

  • Technegas is carbon nanoparticles of 5-200 nm, not a gas; it stays in the lungs for hours. It aggregates on standing, so it is used within 10 minutes.
  • If the crucible volume is exceeded, pertechnetate spills from the crucible and free pertechnetate is inhaled; image quality suffers.
  • DTPA aerosol is droplets of 1-3 µm; its lung clearance half-time is 30-60 minutes, so imaging is done immediately.
  • In children (6 years and older) the activity is reduced by weight; target count rate 500-1000 counts/s (Technegas US label).
  • System and set details vary by product; the manual of the system in use applies.

Common mistakes

  • Overfilling the crucible
  • Exceeding 10 minutes
  • Starting without rehearsing with the patient
  • Detaching the set and releasing residual aerosol into the room
  • Not monitoring oxygen saturation

Safety. Use the filtered set and do not stand close to the patient's exhaled air; dispose of waste in a closed bag; survey the room and surfaces.

Sources: SNMMI/EANM/ACNM/ANZSNM V/Q sintigrafisi prosedür standardı (2026; doi 10.2967/jnmt.126.273453) · Technegas ABD ürün bilgisi (DailyMed 2026) · Farrell ve ark., J Nucl Med Technol 2025;53:11-13 (doi 10.2967/jnmt.125.269536). Last reviewed: 23 November 2026.