앵거 논리(Anger logic)와 결정 두께
광전자 증배관 신호의 가중 평균이 위치를, 그 합이 에너지를 준다. 결정이 두꺼우면 광자를 더 많이 잡지만 빛이 넓게 퍼져 위치가 덜 확실해진다.
제목과 요약은 기계 보조 번역(초안)입니다. 애니메이션은 영어로 되어 있습니다. 영어 전체 페이지 →
In a gamma camera the scintillation light spreads to the PMTs, position is found by a weighted mean and energy by the summed signal; the fraction of photons absorbed and the light spread by crystal thickness are shown.
Key point: the weighted mean of the PMT signals gives the position and their sum the energy; a thick crystal catches more photons but spreads the light wider, so the position becomes less certain.
Source: book (NaI(Tl) 3/8 inch, 3.67 g/cm³; X = Σ(Xᵢ·Sᵢ)/ΣSᵢ, Z = ΣSᵢ; Tc-99m window 126–154 keV). Absorption from the NIST XCOM NaI total attenuation coefficient (coherent included); the book's ~30% at 364 keV is the photopeak efficiency, as part of the interacting photons escape by Compton scattering. In the crystal-thickness step the side view is drawn at 2× vertical scale (light guide 12 mm). Light spot: point light source, reflections ignored, FWHM ≈ 1.53 × distance to the PMT plane. Model assumptions: 37 PMTs, 76 mm pitch, 12 mm light guide, ~930 photoelectrons at 140 keV, Poisson noise; PMTs below 10% of the largest signal ignored (like the threshold in modern cameras); scattered photons ~25% of events, 95–125 keV (illustrative).
How to use: a single click on the drawing pauses and resumes; in the crystal thickness step ↑ ↓ change the thickness, the buttons below change thickness and energy; ← → change steps.