呼吸配准不良(respiratory misregistration)

CT是瞬间快照,PET是呼吸平均;深吸气CT会在肝顶部形成冷带,CT中膈肌高于PET时会在肺底形成热伪影;浅呼吸下CT或门控可避免这一问题。

标题和摘要为机器辅助翻译(草稿);动画内容为英文。 英文完整页面 →

In a coronal section the diaphragm moves with breathing; PET sees the average of minutes of breathing, CT a single moment. The moment of the CT is chosen; the second step shows how the difference causes a cold band at the liver dome or a hot artefact at the lung base in CT attenuation correction.

Continuous stream · no seeking

Key point: CT is a snapshot, PET a breathing average; a CT in deep inspiration gives a cold band at the liver dome, a diaphragm higher than in PET a hot artefact at the lung base; CT in shallow breathing or gating prevents this.

Source: book (CT instantaneous, PET average; cold band at the liver dome or hot artefact at the lung base; solution CT in shallow breathing, respiratory gating; MRAC µ values lung 0.0224, soft tissue 0.100 cm⁻¹). Breathing curve from the Lujan et al. 1999 model (d = A·cos⁴(πt/τ), more time at expiration; mean 3/8·A). Amplitude 2 cm, period 4 s and CT positions (4/2/0 cm, shallow ≈ mean) illustrative.

How to use: a single click on the drawing pauses and resumes; ↑ ↓ or the buttons below change the breathing moment of the CT; ← → change steps.