SPECT假定所有投影彼此一致;患者运动造成断裂和双轮廓,固定的探测器缺陷造成环状伪影,旋转中心漂移使点源模糊或呈环形。
标题和摘要为机器辅助翻译(草稿);动画内容为英文。 英文完整页面 →
Three SPECT artefacts are shown with real projection and filtered back-projection: a sinogram break and double contour with patient motion halfway through acquisition, a sinogram stripe and a ring in the slice with detector non-uniformity, and blurring of a point source that turns into a ring with a centre-of-rotation (COR) error. The size of the error is chosen in each step.
Step 1/3
Patient motion
If the patient shifts sideways halfway through acquisition (at 180°) the projections no longer agree: the sinogram shows a break and the slice a double contour and streaks. Motion is recognised in the projection cine or the sinogram; the remedy is a repeat acquisition or motion correction. Change the shift.
Key point: SPECT assumes all projections are consistent; patient motion gives a break and double contour, a fixed detector defect a ring, and a centre-of-rotation shift a blurred or ring-shaped point source.
Source: book (sinogram break, double contour/streaks; non-uniformity → ring; COR: blurred if small, doughnut if large; COR test every 3-6 months). Calculation: same projector as the tomography animation, 64 × 64 (4 mm pixels), 96 angles / 360°, ramp-filtered FBP with 3 × 3 smoothing, noise-free. Shift, defect and COR values illustrative.
How to use: ↑ ↓ or the buttons below change the size of the fault (shift, sensitivity loss, centre-of-rotation error); the sinogram and the slice are recomputed for each choice; ← → change steps.