断层图像重建
简单反投影(simple backprojection)会产生星状伪影和模糊;斜坡滤波函数可纠正这一点,但会增加噪声;迭代重建逐步使图像与测得的投影相符。
标题和摘要为机器辅助翻译(草稿);动画内容为英文。 英文完整页面 →
Projections and sinogram of a phantom and images reconstructed with simple backprojection, filtered backprojection (FBP) and MLEM/OSEM, computed in the browser.
Step 1/4
Projection and sinogram
The detector rotates around the patient; at each angle the sum of all activity along that direction (line integral) gives a projection profile. Stacking the profiles in angle order gives the sinogram. A point source at the centre is seen in the same place at every angle and draws a straight vertical line; an off-centre point draws a sine curve, hence the name.
The calculations are real: 64 × 64 matrix, projections over 360° (parallel beam), ramp (Ram-Lak) filter, MLEM/OSEM. The phantom has a body, two hot lesions and one cold region. Noisy data have Poisson noise added; total counts are schematic.
Key point: simple backprojection gives a star artefact and blurring; the ramp filter corrects this but raises noise; iterative methods fit the image step by step to the measured projections.
Source: book; 64 × 64 matrix, 360° parallel-beam projections, Ram-Lak filter, MLEM/OSEM (evenly spaced subsets), Poisson noise; total counts schematic.
How to use: a single click on the images pauses and resumes; ↑ ↓ or the buttons above change that step's choice (object, number of projections, number of subsets); ← → change steps.