Leucocyte labelling (Tc-99m HMPAO, In-111 oxine)
Awaiting physician review
To separate the patient's own leucocytes without damage, label them with Tc-99m HMPAO or In-111 oxine and return them aseptically to the same patient.
What you need
- Laminar flow cabinet or cell isolator; sterile gloves, cap, mask
- 60 mL syringe, ACD-A, needle of at least 20G; 10% HES (200 kDa)
- Sterile centrifuge tubes and a centrifuge that can run with the brake off
- Unstabilised HMPAO kit and fresh eluate, or In-111 oxine
- Dose calibrator; trypan blue and haemocytometer (periodic checks)
- Patient identification labels (for every tube and syringe)
Steps
- Identify the patient with two identifiers; do not bring another patient's blood into the cabinet while working with this patient's blood. Label every tube and syringe with the patient's name.
- Draw 9 mL of ACD-A into a 60 mL syringe and slowly draw 51 mL of blood through a needle of at least 20G (no foaming); mix by inversion, do not shake. At least 2 × 10⁸ leucocytes are needed for a good labelling efficiency.
- Centrifuge 15 mL of the blood-ACD mixture at 2000 g for 10 min; separate the cell-free plasma (CFP) on top and keep it for labelling and resuspension.
- Add 4.5 mL of 10% HES to the remaining 45 mL (10:1) and invert gently; stand the syringe nozzle-up for 30-45 min to let the red cells sediment.
- Transfer the leucocyte-rich plasma to a tube with a long needle of at least 20G; centrifuge at 150 g for 5 min with the brake off. Remove the platelet-rich plasma and gently resuspend the leucocyte pellet in 1 mL of CFP.
- Tc-99m HMPAO: prepare an unstabilised kit with fresh eluate and within 20 min add 1 mL (about 750-1000 MBq) to the cells; incubate for 10 min at room temperature with occasional swirling. With In-111 oxine the cells are labelled in a plasma-free medium (saline or PBS).
- Add at least 3 mL (preferably 10 mL) of CFP and centrifuge at 150 g for 5 min; separate the supernatant. Measure the pellet and supernatant to calculate labelling efficiency; resuspend the cells in 3-5 mL of CFP.
- Check for clumps; if clumps do not disperse, do not give the product. Draw up the dose, assay it and inject it slowly through a needle of at least 22G within 1 hour at the latest.
Key points
- Acceptance: labelling efficiency of 40-80% is expected for Tc-99m HMPAO and 50-80% for In-111 oxine; below the lower limit, check for clumping under the microscope and viability with trypan blue. If more than 4% of cells are dead on trypan blue, the product is not given.
- Lipophilic Tc-HMPAO converts over time to a hydrophilic form that cannot enter cells; hence fresh eluate and use within 20 min of preparation. Only unstabilised HMPAO is used for leucocyte labelling.
- In-111 oxine binds to plasma transferrin; no plasma is used in the labelling step and the cells are washed. Labelling efficiency is higher with In-111 but its half-life and energy are higher; it is preferred for assessing the kidneys, bladder and bowel.
- Adult activity: Tc-99m leucocytes 185-370 MBq (EANM 2010 dosimetry table, SNM; the procedure section of the same guideline states 370-740 MBq), In-111 leucocytes 10-18 MBq. Focal lung uptake at 30 minutes suggests cell clumps in the injected sample.
- In leucopenia (neutrophils < 2 × 10³/mm³) an extra syringe of blood may be needed; small blood volumes such as 20 mL lower the efficiency. In children the blood volume and activity are reduced by weight.
Common mistakes
- Working with two patients' blood at the same time
- Fine needles, fast withdrawal or shaking (cell damage)
- Running the centrifuge with the brake on or at high speed
- Using old eluate or HMPAO older than 20 minutes
- Using plasma during In-111 oxine labelling
- Giving a product with clumps that do not disperse, or exceeding 1 hour
Safety. Blood and radioactivity together: wear waterproof gloves, use a shielded syringe and tube shield, take care with needles, and manage spills as radioactive and biological waste.
Sources: de Vries ve ark., Eur J Nucl Med Mol Imaging 2010;37:842-848 (doi 10.1007/s00259-010-1394-4, PMID 20198473) · Roca ve ark., Eur J Nucl Med Mol Imaging 2010;37:835-841 (doi 10.1007/s00259-010-1393-5, PMID 20198474) · Signore ve ark., Eur J Nucl Med Mol Imaging 2018;45:1816-1831 (doi 10.1007/s00259-018-4052-x, PMID 29850929) · SNM lökosit (HMPAO) prosedür kılavuzu 3.0 (2004) · EANM Technologist's Guide Radiopharmacy (2019), Bölüm 9 · European Nuclear Medicine Guide (EANM), lökosit bölümü. Last reviewed: 23 November 2026.