Quality control · 4 of 8

Dose calibrator periodic tests (accuracy, precision, linearity, geometry)

Awaiting physician review

To document that the calibrator keeps the performance established at acceptance, that its measurements are accurate against a traceable standard, and that it is linear over the whole range of use.

What you need

  • Traceable certified sources (e.g. Co-57, Ba-133, Cs-137); at least two radionuclides
  • High-activity Tc-99m (decay method) or a calibrated set of lead sleeves (shield method)
  • Containers used clinically (vials, syringes) and saline for dilution
  • The calibrator's source holder, tongs, gloves, dosimeter
  • Record form or spreadsheet (results kept available for inspection)

Steps

  1. Complete the daily tests (inspection, high voltage, clock, zero, background, constancy); periodic tests follow them.
  2. Accuracy: assay each certified source in the holder on its own radionuclide setting; compare with the decay-corrected certificate value. The difference must not exceed ±5% (±2% for secondary standard and reference calibrators).
  3. Precision (reproducibility): measure a long-lived source of more than 3.7 MBq in the holder 10 times in a row; each reading must be within ±1% of the mean.
  4. Linearity, decay method: assay high-activity Tc-99m in the holder at set intervals (e.g. every 12 hours for three days) until it falls to about 1 MBq. Compare the readings with the expected decay curve on a semi-logarithmic plot; each point must be within ±5%.
  5. Linearity, shield method: place the lead sleeves around the same Tc-99m source in the manufacturer's order and assay with each; compare with the sleeves' baseline calibration (±5%).
  6. Geometry (at acceptance, after repair and after relocation): assay the same activity in the containers and volumes used clinically (starting with a small volume and diluting with saline). If the effect exceeds 5%, determine a correction factor for that container and volume.
  7. Supplier equivalence (at acceptance and annually): compare the supplier's stated activity (decay-corrected) with your own assay.
  8. Record the results. If a limit is exceeded, do not use the calibrator for clinical assays, investigate, call service if needed, and repeat the acceptance tests after repair.

Key points

  • Frequency (AAPM TG-181, EANM 2017): accuracy, precision, linearity and supplier equivalence at acceptance, after repair and annually. For linearity the decay method is used at acceptance and after repair; the shield method is sufficient for the annual test. Some national rules and older sources require linearity quarterly with a ±10% limit; the regulations that apply to the institution take precedence.
  • IAEA TRS 454 (2006) limits: precision ±1%, accuracy ±5% (±2% in secondary standard laboratories), linearity within 5% over the operating range, background ±20% of the mean, constancy ±2% of reference; a new factor if the geometry effect exceeds 5%.
  • A calibrator is linear if the ratio of measured to expected response stays constant over the whole current range; the decay method therefore starts from the highest activity measured clinically (highest current) and goes down to about 1 MBq.
  • The sleeves are calibrated on a calibrator whose decay-method linearity is within ±5%; this is repeated for a new set of sleeves.
  • Target for patient doses: diagnostic dosages within ±10% of the prescription, therapeutic dosages within ±5% where practicable (AAPM TG-181). The daily check is a separate card (doz-kalibratoru-gunluk).

Common mistakes

  • Forgetting decay correction in the accuracy test
  • Measuring the source outside the holder or at a different height
  • Not starting linearity from the highest activity used clinically
  • Using a new set of sleeves without calibrating them
  • Not repeating the geometry test when the container type changes
  • Continuing to use the calibrator after an out-of-limit result

Safety. Carry the high-activity Tc-99m source with tongs and in a shield; keep time near the source short when changing sleeves; store certified sources locked and shielded.

Sources: AAPM Report 181 (TG-181, 2012) · EANM Technologist's Guide: Quality control of NM instrumentation (2017), Bölüm 9 · IAEA TRS 454 (2006), Ek VII · Zanzonico, J Nucl Med 2008;49:1114-1131 (doi 10.2967/jnumed.107.050203, PMID 18587088) · NPL Good Practice Guide 93 (2006). Last reviewed: 23 November 2026.