Quality control · 4 of 8
Dose calibrator periodic tests (accuracy, precision, linearity, geometry)
Awaiting physician review
To document that the calibrator keeps the performance established at acceptance, that its measurements are accurate against a traceable standard, and that it is linear over the whole range of use.
What you need
- Traceable certified sources (e.g. Co-57, Ba-133, Cs-137); at least two radionuclides
- High-activity Tc-99m (decay method) or a calibrated set of lead sleeves (shield method)
- Containers used clinically (vials, syringes) and saline for dilution
- The calibrator's source holder, tongs, gloves, dosimeter
- Record form or spreadsheet (results kept available for inspection)
Steps
- Complete the daily tests (inspection, high voltage, clock, zero, background, constancy); periodic tests follow them.
- Accuracy: assay each certified source in the holder on its own radionuclide setting; compare with the decay-corrected certificate value. The difference must not exceed ±5% (±2% for secondary standard and reference calibrators).
- Precision (reproducibility): measure a long-lived source of more than 3.7 MBq in the holder 10 times in a row; each reading must be within ±1% of the mean.
- Linearity, decay method: assay high-activity Tc-99m in the holder at set intervals (e.g. every 12 hours for three days) until it falls to about 1 MBq. Compare the readings with the expected decay curve on a semi-logarithmic plot; each point must be within ±5%.
- Linearity, shield method: place the lead sleeves around the same Tc-99m source in the manufacturer's order and assay with each; compare with the sleeves' baseline calibration (±5%).
- Geometry (at acceptance, after repair and after relocation): assay the same activity in the containers and volumes used clinically (starting with a small volume and diluting with saline). If the effect exceeds 5%, determine a correction factor for that container and volume.
- Supplier equivalence (at acceptance and annually): compare the supplier's stated activity (decay-corrected) with your own assay.
- Record the results. If a limit is exceeded, do not use the calibrator for clinical assays, investigate, call service if needed, and repeat the acceptance tests after repair.
Key points
- Frequency (AAPM TG-181, EANM 2017): accuracy, precision, linearity and supplier equivalence at acceptance, after repair and annually. For linearity the decay method is used at acceptance and after repair; the shield method is sufficient for the annual test. Some national rules and older sources require linearity quarterly with a ±10% limit; the regulations that apply to the institution take precedence.
- IAEA TRS 454 (2006) limits: precision ±1%, accuracy ±5% (±2% in secondary standard laboratories), linearity within 5% over the operating range, background ±20% of the mean, constancy ±2% of reference; a new factor if the geometry effect exceeds 5%.
- A calibrator is linear if the ratio of measured to expected response stays constant over the whole current range; the decay method therefore starts from the highest activity measured clinically (highest current) and goes down to about 1 MBq.
- The sleeves are calibrated on a calibrator whose decay-method linearity is within ±5%; this is repeated for a new set of sleeves.
- Target for patient doses: diagnostic dosages within ±10% of the prescription, therapeutic dosages within ±5% where practicable (AAPM TG-181). The daily check is a separate card (doz-kalibratoru-gunluk).
Common mistakes
- Forgetting decay correction in the accuracy test
- Measuring the source outside the holder or at a different height
- Not starting linearity from the highest activity used clinically
- Using a new set of sleeves without calibrating them
- Not repeating the geometry test when the container type changes
- Continuing to use the calibrator after an out-of-limit result
Safety. Carry the high-activity Tc-99m source with tongs and in a shield; keep time near the source short when changing sleeves; store certified sources locked and shielded.
Sources: AAPM Report 181 (TG-181, 2012) · EANM Technologist's Guide: Quality control of NM instrumentation (2017), Bölüm 9 · IAEA TRS 454 (2006), Ek VII · Zanzonico, J Nucl Med 2008;49:1114-1131 (doi 10.2967/jnumed.107.050203, PMID 18587088) · NPL Good Practice Guide 93 (2006). Last reviewed: 23 November 2026.