Quality control · 5 of 8

Gamma camera daily check (peak, background, uniformity)

Awaiting physician review

To show before the first patient that the camera's energy window is correct, the background is clean and the detector response is uniform.

What you need

  • Co-57 flood (sheet) source (extrinsic) or a Tc-99m point source of about 18.5 MBq in < 1 mL (intrinsic)
  • The collimators to be used
  • Source holder or a ceiling/floor mark (for the point source)
  • QC record form or system software

Steps

  1. Visual and safety check: collimator damage, loose parts, cables; test the collision sensor (touch pad) and the emergency stop.
  2. Background: with no source present, measure the background count rate in every energy window to be used; if higher than expected, look for contamination of the camera, floor or surroundings or for an unshielded source.
  3. Peak: check the energy spectrum for each radionuclide; the photopeak must be centred in the window (usually 20%, Eγ ± 10%). At least once a day, ideally before each patient.
  4. Uniformity, extrinsic: place the Co-57 sheet source on the collimator. Intrinsic: remove the collimator and place the Tc-99m point source centred over the detector at least 3, preferably 5, times its largest dimension away; the count rate should not exceed 25 000 counts/s.
  5. Acquire 10-15 million counts (a lower-count daily image may be used on a large-field camera; follow the product and institutional method).
  6. Inspect the image visually (PMT pattern, cold or hot areas, edges); calculate integral (IU) and differential (DU) uniformity in the central (CFOV) and useful (UFOV) fields of view with the software.
  7. If IU or DU exceeds 5% or there is a visible defect, do not image patients: repeat the peaking, and if the problem persists acquire a new high-count (60-100 million) uniformity correction table or call service.
  8. Record the results with date, time, source and operator; track the values over time on a graph.

Key points

  • If a daily low-count intrinsic test is chosen, each collimator is checked weekly or monthly with a high-count extrinsic test (EANM 2017). On multi-crystal (semiconductor) cameras only the extrinsic image is meaningful (AAPM 177).
  • IU is usually 3% or better on modern cameras. Causes of non-uniformity: an outdated correction table, wrong peaking, a PMT uncoupled from the crystal, a cracked crystal, and switched-off or corrupted correction tables.
  • The crystal can crack with temperature changes faster than about 5 °C per hour; room temperature must not change suddenly.
  • In SPECT, non-uniformity causes a ring (bull's-eye) artefact; IU ≈ 10% produced an obvious ring.
  • Frequencies and limits are general recommendations; the manufacturer's and the institution's QC programme apply.

Common mistakes

  • Using the intrinsic point source with the collimator fitted
  • Placing the point source too close (non-uniform flux)
  • Running too high a count rate (dead time)
  • Imaging with the peak set for the wrong radionuclide
  • Imaging patients after a failed daily test

Safety. Store the Co-57 source shielded and position it quickly; carry the Tc-99m point source with tongs and discard it as waste after use.

Sources: Zanzonico, J Nucl Med 2008;49:1114-1131 (doi 10.2967/jnumed.107.050203, PMID 18587088) · EANM Technologist's Guide: Quality control of NM instrumentation (2017) · European Nuclear Medicine Guide (EANM), Bölüm 21 kalite kontrol · AAPM Report 177 (2019), gama kamera ve SPECT kabul ve yıllık testleri. Last reviewed: 23 November 2026.