Quality control · 6 of 8

Gamma camera and SPECT periodic tests (resolution, linearity, COR, tomographic performance)

Awaiting physician review

To document at regular intervals the camera's spatial resolution and linearity, the SPECT centre-of-rotation alignment and tomographic image quality.

What you need

  • Four-quadrant bar phantom (2, 2.5, 3 and 4 mm)
  • Tc-99m point source
  • Tomographic phantom (e.g. Jaszczak type, cold spheres and rods)
  • SPECT/CT alignment phantom (manufacturer's)
  • Collimators and Co-57 source for high-count extrinsic uniformity

Steps

  1. Resolution and linearity (weekly to monthly): place the bar phantom on the uncollimated detector, position the Tc-99m point source at least 5 detector dimensions away, acquire 5-10 million counts.
  2. At least the 3 and 4 mm bars must be resolved and part of the 2.5 mm bars visible; all bars must be straight. Rotate the phantom through four orientations to test every region.
  3. COR (weekly to monthly; per manufacturer): acquire a 360° SPECT of a point source by the manufacturer's method; the software calculates the deviation of the x position from the sine curve and updates the correction table.
  4. Collimator uniformity: if a daily intrinsic test is used, acquire a high-count extrinsic image for each collimator (weekly or monthly).
  5. SPECT/CT alignment (monthly or per manufacturer): scan the phantom and check that the SPECT and CT images coincide.
  6. Tomographic performance (quarterly): fill the tomographic phantom with about 370 MBq of Tc-99m; acquire at least 64 projections at 40 s or more per projection. There must be no ring artefact in the uniform section; cold spheres and rods are compared with previous results.
  7. Record the results; compare them with previous values and take the camera out of clinical use and call service if out of limits.
  8. The annual physics survey and acceptance tests (NEMA methods) are done with the medical physicist and service; the technologist files the results.

Key points

  • Frequencies vary by source: resolution and linearity weekly (Zanzonico 2008) or monthly (European Nuclear Medicine Guide); COR weekly or monthly; tomographic phantom quarterly. The manufacturer's and institution's programme apply.
  • A COR misalignment as small as 3.2 mm (half a pixel in a 64 × 64 matrix) produces perceptible blurring in SPECT images; 6.4 mm (one pixel) causes marked degradation.
  • The linearity correction table rarely changes after installation and is updated by service; if switched off, the PMT pattern becomes obvious and IU approaches 20%.
  • On multi-head cameras, head-to-head sensitivity differences and head tilt are checked periodically.
  • The daily check is a separate card (gama-kamera-gunluk).

Common mistakes

  • Testing the bar phantom in only one orientation
  • Checking COR less often than the manufacturer recommends
  • Imaging the phantom with too few counts and blaming artefacts on noise
  • Not comparing with previous results (missing slow degradation)

Safety. Transfer the Tc-99m for the phantom with a shielded syringe, fill the phantom over a tray to avoid spills, and survey the area at the end.

Sources: Zanzonico, J Nucl Med 2008;49:1114-1131 (doi 10.2967/jnumed.107.050203, PMID 18587088) · European Nuclear Medicine Guide (EANM), Bölüm 21 kalite kontrol · EANM Technologist's Guide: Quality control of NM instrumentation (2017) · AAPM Report 177 (2019) · ANZSNM gama kamera en az QC çizelgesi (2013; kafalar arası duyarlılık, kafa eğimi). Last reviewed: 23 November 2026.