Quality control · 6 of 8
Gamma camera and SPECT periodic tests (resolution, linearity, COR, tomographic performance)
Awaiting physician review
To document at regular intervals the camera's spatial resolution and linearity, the SPECT centre-of-rotation alignment and tomographic image quality.
What you need
- Four-quadrant bar phantom (2, 2.5, 3 and 4 mm)
- Tc-99m point source
- Tomographic phantom (e.g. Jaszczak type, cold spheres and rods)
- SPECT/CT alignment phantom (manufacturer's)
- Collimators and Co-57 source for high-count extrinsic uniformity
Steps
- Resolution and linearity (weekly to monthly): place the bar phantom on the uncollimated detector, position the Tc-99m point source at least 5 detector dimensions away, acquire 5-10 million counts.
- At least the 3 and 4 mm bars must be resolved and part of the 2.5 mm bars visible; all bars must be straight. Rotate the phantom through four orientations to test every region.
- COR (weekly to monthly; per manufacturer): acquire a 360° SPECT of a point source by the manufacturer's method; the software calculates the deviation of the x position from the sine curve and updates the correction table.
- Collimator uniformity: if a daily intrinsic test is used, acquire a high-count extrinsic image for each collimator (weekly or monthly).
- SPECT/CT alignment (monthly or per manufacturer): scan the phantom and check that the SPECT and CT images coincide.
- Tomographic performance (quarterly): fill the tomographic phantom with about 370 MBq of Tc-99m; acquire at least 64 projections at 40 s or more per projection. There must be no ring artefact in the uniform section; cold spheres and rods are compared with previous results.
- Record the results; compare them with previous values and take the camera out of clinical use and call service if out of limits.
- The annual physics survey and acceptance tests (NEMA methods) are done with the medical physicist and service; the technologist files the results.
Key points
- Frequencies vary by source: resolution and linearity weekly (Zanzonico 2008) or monthly (European Nuclear Medicine Guide); COR weekly or monthly; tomographic phantom quarterly. The manufacturer's and institution's programme apply.
- A COR misalignment as small as 3.2 mm (half a pixel in a 64 × 64 matrix) produces perceptible blurring in SPECT images; 6.4 mm (one pixel) causes marked degradation.
- The linearity correction table rarely changes after installation and is updated by service; if switched off, the PMT pattern becomes obvious and IU approaches 20%.
- On multi-head cameras, head-to-head sensitivity differences and head tilt are checked periodically.
- The daily check is a separate card (gama-kamera-gunluk).
Common mistakes
- Testing the bar phantom in only one orientation
- Checking COR less often than the manufacturer recommends
- Imaging the phantom with too few counts and blaming artefacts on noise
- Not comparing with previous results (missing slow degradation)
Safety. Transfer the Tc-99m for the phantom with a shielded syringe, fill the phantom over a tray to avoid spills, and survey the area at the end.
Sources: Zanzonico, J Nucl Med 2008;49:1114-1131 (doi 10.2967/jnumed.107.050203, PMID 18587088) · European Nuclear Medicine Guide (EANM), Bölüm 21 kalite kontrol · EANM Technologist's Guide: Quality control of NM instrumentation (2017) · AAPM Report 177 (2019) · ANZSNM gama kamera en az QC çizelgesi (2013; kafalar arası duyarlılık, kafa eğimi). Last reviewed: 23 November 2026.