Quality control · 7 of 8

PET and PET/CT daily quality control and periodic calibrations

Awaiting physician review

To show before the first patient that the PET detectors and electronics are within the manufacturer's limits, and to renew the normalisation and activity calibration on time.

What you need

  • Manufacturer's daily QC source or phantom (e.g. Ge-68 cylinder or rod source)
  • Source holder
  • Water phantom for CT
  • Manufacturer's phantom and F-18 for normalisation and activity calibration (periodic)
  • Record form or system report

Steps

  1. Start the system; warm up the CT tube with the manufacturer's procedure.
  2. Run the daily PET check as the manufacturer describes: place the source in its holder and start the routine (usually under 30 minutes). The system tests gains, energy window, timing, detector status and calibration factor.
  3. Read the report: green means compliant, yellow a mild and red a marked deviation. Inspect the sinogram: a blank diagonal line indicates a faulty detector element, a diagonal band a faulty detector block.
  4. If there is a fault, do not image patients; apply the manufacturer's recommended correction (e.g. gain adjustment), and call service if it persists.
  5. CT daily test: scan the water phantom with clinical parameters; the CT number difference between centre and peripheral regions must not exceed 5 HU, and the CT number of water and noise must be within the manufacturer's limits.
  6. Tomographic uniformity: with a Ge-68 cylinder daily or at least weekly; there must be no ring or band artefact.
  7. Normalisation and activity (well-counter) calibration: at installation, after major service, if image quality deteriorates, and at least annually.
  8. Record the results; for calibrations that affect SUV, check that cross-calibration with the dose calibrator is current.

Key points

  • The daily check is manufacturer-specific and runs automatically or semi-automatically; the user follows the manufacturer's recommendation exactly (EFOMP).
  • The sinogram shows a faulty detector better than the reconstructed image: a faulty block makes an obvious band in the sinogram but may not be noticeable in the image.
  • Normalisation is the table that corrects sensitivity differences between thousands of detector elements (the counterpart of uniformity correction in a gamma camera).
  • The activity (well-counter) calibration converts voxel counts to kBq/mL; if it drifts, SUV is directly affected.
  • Frequencies are general recommendations; the manufacturer's and institution's programme apply.

Common mistakes

  • Skipping the daily check and imaging patients
  • Leaving a yellow warning without follow-up
  • Reading only the report and not the sinogram
  • Leaving the Ge-68 source unshielded
  • Postponing normalisation beyond the annual interval

Safety. Ge-68 sources are long-lived (T½ 271 days): store them shielded and locked, carry them with tongs and keep your distance when positioning them.

Sources: Zanzonico, J Nucl Med 2008;49:1114-1131 (doi 10.2967/jnumed.107.050203, PMID 18587088) · EFOMP kılavuzu: PET/BT ve PET/MR kalite kontrolleri · EANM Technologist's Guide: Quality control of NM instrumentation (2017). Last reviewed: 23 November 2026.