Quality control · 8 of 8
Well counter and thyroid uptake probe check (peak, background, constancy, efficiency, chi-square)
Awaiting physician review
To show that sample and patient counts are made in the correct energy window, with a clean background and a known efficiency.
What you need
- Traceable long-lived sources (e.g. Co-57, Ba-133, Cs-137)
- Empty counting tube (for background)
- Calibrated sample (about 37 kBq in ~1 mL; for efficiency)
- Neck phantom and capsule/standard for the thyroid probe
- Record form
Steps
- Peak: check the spectrum for each radionuclide to be used; the photopeak must be centred in the 20% window (Eγ ± 10%).
- Background: count an empty tube daily in each window to be used; if high, look for contamination of the well or a nearby source. Always count a blank tube with sample counts.
- Constancy: count at least one traceable source daily; the background-subtracted count rate must stay within ±10% from day to day.
- Energy resolution (at least quarterly): measure the full width at half maximum (FWHM) of the photopeak as a percentage with Co-57.
- Chi-square test (at least quarterly): count the same source repeatedly (e.g. 10 times) and calculate χ²; P should be between 0.05 and 0.95, otherwise suspect a counter fault.
- Efficiency (at installation and annually; for the thyroid probe for each patient): count a sample of known activity or a standard in the phantom; efficiency = (gross − background) / decay-corrected activity.
- Thyroid probe: patient and standard are counted in the same geometry (neck phantom, same distance); background and counting time follow the institution's uptake protocol.
- Record the results; do not use the instrument after an out-of-limit result and investigate.
Key points
- The well counter is very sensitive and counts reliably only up to about 37 kBq; at higher activities dead-time losses distort the measurement. It is for low activities the dose calibrator cannot measure (wipe tests, blood, urine).
- Minimum detectable activity: the gross count rate must exceed background by at least 3 standard deviations.
- If the source and the clinical radionuclide differ in energy, so does efficiency; at installation the radionuclide and its surrogate are counted together to obtain a conversion factor, after which annual checks can use only the surrogate (e.g. Co-57 for Tc-99m).
- For the intraoperative probe, battery, bias voltage, background and constancy (±10%) are checked daily; the source is counted in the same geometry (cap) every day.
- Frequencies are general recommendations; the manufacturer's and institution's programme apply.
Common mistakes
- Counting samples without a blank tube
- Counting a high-activity sample in the well (dead time)
- Counting with the wrong window
- Counting patient and standard at different distances with the thyroid probe
- Never evaluating the chi-square result
Safety. Store sources shielded; place sample tubes closed and clean on the outside; if background rises, look for contamination of the well.
Sources: Zanzonico, J Nucl Med 2008;49:1114-1131 (doi 10.2967/jnumed.107.050203, PMID 18587088) · EANM Technologist's Guide: Quality control of NM instrumentation (2017), Bölüm 11. Last reviewed: 23 November 2026.