Drawing up and assaying a dose with a syringe shield
Awaiting physician review
To draw a patient dose aseptically from a multidose vial into a shielded syringe, assay it correctly, label it and keep hand dose low.
What you need
- Shielded vial with the labelled radiopharmaceutical
- Syringe shield (lead or tungsten for Tc-99m; tungsten for PET)
- Single-use sterile syringe and needle, alcohol swab
- Dose calibrator, patient label, record form
- Gloves, ring dosimeter
Steps
- Check the request and the patient: radiopharmaceutical, activity (MBq), route and time.
- Calculate the volume needed from the vial concentration and the elapsed time.
- Swab the vial stopper with alcohol and let it dry. Fit the syringe into its shield; without taking the vial out of its shield, draw the dose slowly, and do not inject air into the vial (for oxidation-sensitive kits).
- Remove the needle and close the syringe with a cap or a new sterile needle; check the outside for contamination.
- Assay the syringe in the dose calibrator on the correct radionuclide setting and in the correct geometry.
- If the measured activity differs from the request, adjust it: separate the excess or add what is missing; re-assay.
- Label the syringe shield with patient name, radiopharmaceutical, activity, assay time and volume.
- After injection, assay and record the residual activity in the syringe; administered activity = before − after.
Key points
- Target: diagnostic dosages within ±10% of the prescription, therapeutic dosages within ±5% where practicable (AAPM TG-181).
- A syringe shield and an automatic dispenser reduced finger dose by more than a factor of 10 in the procedures measured (Montgomery et al. 1999). In PET, fingertip doses from manual dispensing and injection can exceed the annual limit even with a syringe shield (Covens et al. 2010).
- No air is introduced into multidose kit vials; if tin is oxidised, free pertechnetate rises.
- F-18 multidose vial details are on a separate card (f18-dagitim).
- Patient identity and syringe label are matched at every step; an unlabelled syringe is never used.
Common mistakes
- Drawing or carrying the syringe unshielded
- Wrong radionuclide setting on the calibrator
- Unlabelled syringe
- Not assaying the residual
- Injecting air into the vial
Safety. Use a shielded syringe and tongs; keep the vial in its shield; keep your fingers away from the unshielded end of the syringe; wear the ring dosimeter on the dominant hand, on the palm side.
Sources: AAPM Report 181 (TG-181, 2012) · Montgomery ve ark., Nucl Med Commun 1999;20:189-194 (doi 10.1097/00006231-199902000-00013, PMID 10088170) · Covens ve ark., Radiat Prot Dosimetry 2010;140:250-258 (doi 10.1093/rpd/ncq110, PMID 20335185) · EANM kemik sintigrafisi kılavuzu (Eur J Nucl Med Mol Imaging 2016; doi 10.1007/s00259-016-3415-4, PMID 27262701), radyasyondan korunma. Last reviewed: 23 November 2026.