Patient procedures · 2 of 7

Drawing up and assaying a dose with a syringe shield

Awaiting physician review

To draw a patient dose aseptically from a multidose vial into a shielded syringe, assay it correctly, label it and keep hand dose low.

What you need

  • Shielded vial with the labelled radiopharmaceutical
  • Syringe shield (lead or tungsten for Tc-99m; tungsten for PET)
  • Single-use sterile syringe and needle, alcohol swab
  • Dose calibrator, patient label, record form
  • Gloves, ring dosimeter

Steps

  1. Check the request and the patient: radiopharmaceutical, activity (MBq), route and time.
  2. Calculate the volume needed from the vial concentration and the elapsed time.
  3. Swab the vial stopper with alcohol and let it dry. Fit the syringe into its shield; without taking the vial out of its shield, draw the dose slowly, and do not inject air into the vial (for oxidation-sensitive kits).
  4. Remove the needle and close the syringe with a cap or a new sterile needle; check the outside for contamination.
  5. Assay the syringe in the dose calibrator on the correct radionuclide setting and in the correct geometry.
  6. If the measured activity differs from the request, adjust it: separate the excess or add what is missing; re-assay.
  7. Label the syringe shield with patient name, radiopharmaceutical, activity, assay time and volume.
  8. After injection, assay and record the residual activity in the syringe; administered activity = before − after.

Key points

  • Target: diagnostic dosages within ±10% of the prescription, therapeutic dosages within ±5% where practicable (AAPM TG-181).
  • A syringe shield and an automatic dispenser reduced finger dose by more than a factor of 10 in the procedures measured (Montgomery et al. 1999). In PET, fingertip doses from manual dispensing and injection can exceed the annual limit even with a syringe shield (Covens et al. 2010).
  • No air is introduced into multidose kit vials; if tin is oxidised, free pertechnetate rises.
  • F-18 multidose vial details are on a separate card (f18-dagitim).
  • Patient identity and syringe label are matched at every step; an unlabelled syringe is never used.

Common mistakes

  • Drawing or carrying the syringe unshielded
  • Wrong radionuclide setting on the calibrator
  • Unlabelled syringe
  • Not assaying the residual
  • Injecting air into the vial

Safety. Use a shielded syringe and tongs; keep the vial in its shield; keep your fingers away from the unshielded end of the syringe; wear the ring dosimeter on the dominant hand, on the palm side.

Sources: AAPM Report 181 (TG-181, 2012) · Montgomery ve ark., Nucl Med Commun 1999;20:189-194 (doi 10.1097/00006231-199902000-00013, PMID 10088170) · Covens ve ark., Radiat Prot Dosimetry 2010;140:250-258 (doi 10.1093/rpd/ncq110, PMID 20335185) · EANM kemik sintigrafisi kılavuzu (Eur J Nucl Med Mol Imaging 2016; doi 10.1007/s00259-016-3415-4, PMID 27262701), radyasyondan korunma. Last reviewed: 23 November 2026.