F-18 dose dispensing (multidose vial, automated dispenser)
Awaiting physician review
To dispense a patient-specific dose from a multidose F-18 vial (e.g. FDG) aseptically, with the correct activity and a low hand dose.
What you need
- Multidose vial and product information (calibration activity, volume, expiry)
- Tungsten syringe shield, thick vial shield or a workstation with a lead glass screen
- Single-use sterile syringe and needle, alcohol swab; 0.9% NaCl if dilution is needed
- Dose calibrator (F-18 setting)
- Automated dispenser or dispenser-injector if available (authorised system)
- Gloves, lab coat, body and ring dosimeters; record form
Steps
- Receive the vial: check package integrity and contamination; record the calibration activity, time, volume and expiry time from the label. Do not use the product if the package is compromised.
- Assay the vial in its shield in the dose calibrator and compare the first reading with the label. Inspect the solution from behind lead glass (clear, free of particles).
- Calculate the volume needed: the required activity divided by the radioactive concentration at that time. The concentration falls as F-18 decays (T½ about 110 min; about 68% of the starting value after 1 hour), so the volume is recalculated for every dose.
- For manual dispensing, swab the stopper with alcohol and let it dry; withdraw the dose through the stopper with a shielded single-use syringe and a sterile needle. Do not open the vial or remove the stopper.
- Assay the syringe in its shield in the dose calibrator; adjust by separating excess without returning it to the vial, or by adding what is missing. If the product information allows, dilute with 0.9% NaCl.
- Label it: patient, radiopharmaceutical, activity, assay time, volume. Re-assay the dose immediately before injection (the product information requires this).
- With an automated dispenser, load the vial as the manufacturer describes and request the patient dose from the system; record the activity it measures and compare the first doses with the dose calibrator.
- At the end of the day, assay and record what is left in the vial and the waste; survey the work area and your hands for contamination.
Key points
- Shelf life differs between products and the product information applies: for one FDG product 14 hours from the end of manufacture and 12 hours after first withdrawal; for another 14 hours from calibration and 8 hours after first use; for a US product 12 hours from the end of synthesis.
- The product information requires the dose to be withdrawn through the stopper with a shielded single-use syringe or an authorised automated system. The vial contains no preservative; each dose is prepared aseptically.
- Because of the 511 keV photons, hand doses are high in PET: even with a syringe shield, fingertip doses from manual dispensing and injection can exceed the annual limit. Automated dispensing and injection reduced hand dose by more than 95%, and automated dispensing alone by up to 80%.
- The accuracy of an automated system is checked on site: in one study two devices were within ±5% and one deviated by up to 30% (at high concentration and small volume). The system's measurement is compared regularly with the dose calibrator.
- Calculator: volume = required activity ÷ [calibration concentration × 0.5^(elapsed time / 110 min)].
Common mistakes
- Assuming a constant concentration and calculating the volume from the morning value
- Piercing the stopper before the alcohol has dried, or opening the vial
- Withdrawing with an unshielded syringe
- Not re-assaying the dose immediately before injection
- Never checking the automated system's measurement
Safety. Use a tungsten syringe shield and a thick vial shield, carry with tongs and keep hands-on time short; wear the ring dosimeter on the hand that receives the most dose.
Sources: SmPC: Fludeoxyglucose (18F)-Curium (2023), EFDEGE (2023), Alliance Medical FDG (UK 2025); EMA çekirdek SmPC FDG (EMA/CHMP/448228/2012); ABD FDG etiketi (DailyMed 2026) · Covens ve ark., Radiat Prot Dosimetry 2010;140:250-258 (doi 10.1093/rpd/ncq110, PMID 20335185) · O'Doherty ve ark., Nucl Med Commun 2014;35:151-159 (doi 10.1097/MNM.0000000000000027, PMID 24217428) · Miyaji ve ark., J Nucl Med Technol 2017;45:91-95 (doi 10.2967/jnmt.116.188243, PMID 28280125). Last reviewed: 23 November 2026.