Radiopharmacy · 4 of 8

Preparation of Lu-177 and Ac-225 therapy doses (receipt, assay, dose withdrawal, records)

Awaiting physician review

To prepare a ready-to-use Lu-177 therapy product, or an Ac-225 dose prepared in-house, for administration with the correct activity, aseptically and with low staff dose.

What you need

  • Product information and the patient-specific dose document
  • Dose calibrator (Lu-177 setting; for Ac-225 a validated calibration factor)
  • Lead glass screen, tongs, shielded syringe; needles described by the product (Pluvicto syringe method: 9 cm 18G long needle, 2.5 cm 20G filtered short venting needle)
  • Contamination monitor and wipe test materials
  • Glovebox for Ac-225
  • Gloves, lab coat, body and ring dosimeters; record form

Steps

  1. Receive the package: check for damage and contamination; compare the product, patient, calibration activity, calibration date and time and expiry time on the label with the dose document.
  2. Check the expiry: Lutathera 72 hours and Pluvicto 120 hours (5 days; US label) after calibration. Calculate the activity at the time of administration by decay (Lu-177 T½ 6.647 days); the product arrives with a volume that delivers 7.4 GBq ± 10% at that time.
  3. Inspect the vial behind a lead glass screen; do not use it if particles or discoloration are present.
  4. Place the vial in the dose calibrator with tongs and assay it on the appropriate setting; record the reading and time.
  5. If the full dose is to be given by the gravity or peristaltic pump method, it is infused from the product's own vial and no separate dose is drawn. For the syringe method, draw the dose into a shielded syringe with the long needle; the venting needle must not touch the solution.
  6. For a reduced dose use the syringe or pump method. If the gravity method is chosen, the dose is adjusted beforehand by withdrawing the excess activity from the vial; otherwise the wrong volume is given.
  7. After administration assay the vial, syringe and set again. The administered activity is the pre-administration reading minus the residual; record it.
  8. Assay an Ac-225 dose only after equilibrium with the gamma-emitting daughters (about 30 min for Fr-221) and with a calibration factor validated for that measurement set-up. Prepare it in a glovebox and wipe-test everything that leaves the box.

Key points

  • Ready-to-use products: Pluvicto 1000 MBq/mL at calibration, 7.5-12.5 mL; Lutathera 370 MBq/mL, 20.5-25 mL. In both, the volume is set to deliver 7.4 GBq at the time of administration. Lutathera is not given as a bolus; the amino acid infusion for renal protection starts 30 min before.
  • The alpha emission of Ac-225 is not measured directly in a dose calibrator; the assay relies on the gammas of its daughters (Fr-221, Bi-213) and is correct only after equilibrium. NIST found calibration settings of 729-735 (×5) for two common calibrator models; the frequently used setting of 775 reads 4.4-5.4% high. The setting depends on the instrument and geometry; cross-validation with a gamma counter or HPGe measurement is recommended.
  • Reading Ac-225 RCP (radio-TLC) also depends on equilibrium: full equilibrium takes about 20 hours. In one study, RCP > 90% measured 2 hours after plate development predicted a true RCP > 97%.
  • Ac-225 is highly radiotoxic: activities used for quality control are small (about 10 kBq outside a fume hood at one centre), preparation is done in a glovebox and the institution's approved procedure is followed.
  • Administration, set-up and extravasation management are on separate cards (lu177-uygulama, enjeksiyon-kacis).

Common mistakes

  • Preparing without checking the expiry time and patient match
  • Giving a reduced dose by the gravity method without adjusting it beforehand
  • Letting the venting needle touch the solution
  • Assaying Ac-225 before equilibrium or with an unvalidated calibrator setting
  • Not assaying the residual activity after administration (the administered dose is then unknown)

Safety. For Lu-177 use a shielded syringe, tongs and lead glass; keep hands-on time short. For Ac-225 external exposure is low but the risk of contamination and intake is high: a glovebox, wipe tests and contamination monitoring are mandatory.

Sources: SmPC/prospektüs: Pluvicto (EMA 2026; ABD 2026), Lutathera (EMA 2026; Kanada 2026; Singapur 2026) · IAEA-TECDOC-2057 (2024), Production and Quality Control of Actinium-225 Radiopharmaceuticals (doi 10.61092/iaea.95h3-2ji2) · Bergeron ve ark., Appl Radiat Isot 2025;217:111630 (doi 10.1016/j.apradiso.2024.111630, PMID 39681031) · Hooijman ve ark., EJNMMI Radiopharm Chem 2024;9:9 (doi 10.1186/s41181-024-00239-1, PMID 38319526) · Kelly ve ark., EJNMMI Radiopharm Chem 2021;6:38 (doi 10.1186/s41181-021-00151-y, PMID 34928478). Last reviewed: 23 November 2026.