Radiopharmacy · 3 of 8

Ga-68 generator elution and kit labelling (PSMA-11, DOTATOC/DOTATATE)

Awaiting physician review

To obtain gallium (Ga-68) chloride eluate meeting Ph. Eur. 2464 from a Ge-68/Ga-68 generator and to prepare a PSMA or somatostatin analogue radiopharmaceutical with an approved kit under aseptic conditions.

What you need

  • Shielded Ge-68/Ga-68 generator and the manufacturer's sterile ultrapure 0.1 M HCl
  • Kit and its product information; buffer vial and, for some kits, a sterile evacuated reaction vial
  • Sterile venting filter (0.2 µm), needles recommended by the product
  • For kits that need heating, a dry heating block and a calibrated thermometer
  • Vial shield, lead glass screen, tongs; gloves, body and ring dosimeters
  • Dose calibrator (Ga-68 setting), pH strips, ITLC-SG strips, 1 M ammonium acetate:methanol (1:1), radio-TLC scanner
  • Record form, label

Steps

  1. Check the time of the last elution in the generator log. If the generator has not been eluted for a long time (GalliaPharm 96 hours or more; Galliad 72 hours or more), pre-elute it as the manufacturer describes and discard the eluate; for GalliaPharm, full yield needs at least 7 hours after the pre-elution.
  2. Prepare the kit as the product information specifies: a refrigerated kit is brought to room temperature; for a kit that needs heating (SomaKit TOC) the dry block is set to 95 °C and the temperature checked with a thermometer. Swab the stoppers with alcohol and let them dry.
  3. Place the vial in its shield and insert the needle with the venting filter. Elute the generator with the volume the product specifies (5 mL for one product, 1.1 mL of 0.1 M HCl for another); depending on the kit, the eluate flows directly into the kit vial (Locametz, SomaKit TOC) or into a separate evacuated reaction vial (Illuccix).
  4. Add the buffer in the order the product information gives (before the eluate for some kits, immediately after it for others) and mix by inverting the vial inside the shield.
  5. Reaction: PSMA-11 kits at room temperature (Locametz EU: 20-30 °C, upright for 5 min); SomaKit TOC at 95 °C for at least 7 and at most 10 min without stirring, then about 10 min cooling in the shield.
  6. Assay the vial in the dose calibrator on the Ga-68 setting; write the activity, assay time, batch number and expiry time on the shield label.
  7. Perform quality control behind a lead glass screen: appearance, pH and the fraction of free and colloidal Ga-68 by ITLC. If a result is outside the product's limit, do not use the product.
  8. Record: eluate activity (yield decay-corrected to the start of elution), QC results, date of the last Ge-68 breakthrough test. Assay each patient dose in the calibrator immediately before administration.

Key points

  • Ga-68 (T½ 67.7 min, positron emitter) is formed by decay of Ge-68 (T½ 271 days) bound to the column and is washed off with dilute HCl. At equilibrium more than 60% of the Ga-68 on the column can be eluted; the yield falls as Ge-68 decays and halves in about 9 months.
  • Under Ph. Eur. 2464, Ge-68 breakthrough must not exceed 0.001% of the eluted Ga-68 activity (decay-corrected). The limit can be exceeded if the generator is not eluted for days, so it is eluted regularly (EMA core SmPC: at least once per working day) and pre-eluted when needed. The test compares the Ga-68 and Ge-68 activities; Ge-68 is measured in a sample after the Ga-68 has decayed.
  • Acceptance limits differ by product and region. Locametz: in the EU non-complexed Ga-68 at most 3%, pH 3.2-6.5, 6 hours; in the US RCP at least 95%. Illuccix: in the EU at most 3%, pH 4.0-5.0, 2 hours after labelling; in the US RCP at least 95%, 4 hours. SomaKit TOC (EU): pH 3.2-3.8, colloidal at most 3%, free at most 2%, 4 hours. NETSPOT (US): RCP at least 95%, pH 3.2-3.8. The product information of the product in use applies.
  • The dry block's set point is not the temperature inside the vial: at one centre, with a 98 °C set point, the vial reached 90.6 °C after 8 minutes. In the same study the needle type (metal contamination) and contact of HCl with the vial stopper lowered RCP. With PSMA-11 kits, non-siliconised needles led to discoloration and QC failure, and not inverting the vial led to inconsistent labelling.
  • Preparation with an automated synthesis module (cassette) is outside the scope of this card; the eluate and the product meet the same pharmacopoeial requirements.

Common mistakes

  • Using the eluate of a generator that has not been eluted for a long time without a pre-elution
  • Not checking the dry block temperature, or exceeding the heating time (SomaKit TOC: at most 10 min)
  • Using needles the product does not recommend (metal contamination, discoloration)
  • Not mixing the eluate and buffer by inversion
  • Diluting the labelled product when the product information does not allow it (prohibited for SomaKit TOC)

Safety. The 511 keV annihilation photons are far more penetrating than Tc-99m photons: use thick lead or tungsten shielding, a lead glass screen and tongs; keep hands-on time with the generator and hot vial short.

Sources: SmPC: GalliaPharm (EMA), Galliad (HPRA 2026) · EMA çekirdek SmPC, 68Ge/68Ga jeneratör (EMA/CHMP/337681/2016) · SmPC/prospektüs: Locametz (EMA 2026; ABD 2025), Illuccix (HPRA 2026; ABD 2025), SomaKit TOC (EMA 2026), NETSPOT (ABD) · Ph. Eur. 2464 (ürün bilgilerinin aktardığı sınır) · Fang ve ark., J Nucl Med Technol 2020;48:263-268 (doi 10.2967/jnmt.119.241224, PMID 32277061) · Wang ve ark., EJNMMI Radiopharm Chem 2024;9:87 (doi 10.1186/s41181-024-00317-4, PMID 39692998) · Nelson ve ark., EJNMMI Radiopharm Chem 2022;7:27 (doi 10.1186/s41181-022-00180-1, PMID 36271969). Last reviewed: 23 November 2026.